Gilson MA-26X Operating Manual page 6

Aqua-check moisture tester
Hide thumbs Also See for MA-26X:

Advertisement

STEP
4. Place the prepared and weighed sample into the Aqua-Check Vessel. For cohesive samples
4
as described above, also place the two Steel Pulverizing Balls into the Vessel.
STEP
5. Using the Long-Handled reagent Scoop, add
5
at least two full scoops of calcium carbide reagent to
the Chamber Cap.
NOTE:
Make sure you have read and
understand the Material Safety Data Sheet
(MSDS) included with each container of
calcium carbide reagent.
6
STEP 6. Holding the Aqua-Check Vessel horizontally, place the chamber cap on the end of the chamber.
Position the black dot on the cap so that it will be facing away from your body when the pressure is
released. Move the stirrup into place over the cap and tighten the top screw to seal the chamber. Use
care during this step not to mix the reagent and sample before the chamber is sealed.
STEP 6: Placing chamber cap on the end of
Gilson Aqua-Check Moisture Tester, MA-26X
STEP 4: Placing sample into Vessel
the chamber
STEP 4: Placing Steel Pulverizing
Balls into Vessel
STEP 5: Adding calcium carbide reagent to
chamber cap
STEP 6: Tightening the top screw to seal
the chamber