PiPetman Fixed is in accordance with isO 8655 standard. 2 - SpEcIfIcatIonS PiPetman Fixed is a high quality pipette that offers excellent accuracy and precision. the figures given in the “Gilson maximum Permissible errors” table were obtained using “PiPetman diamOnd tips”.
Gilson Quality system. Gilson declares that its manufactured pipettes com- ply with the requirements of the isO 8655 standard, by type testing. the adjustement is carried out under strictly defined and monitored conditions (isO 8655-6): • Basis of adjustment, Ex.
4 - uSEr adjuStmEntS PiPetman Fixed is factory calibrated using distilled water and very high precision balances. the nominal value of the PiPetman Fixed may be slightly adjusted to compensate for liquids of different density or viscosity. Adjustment of the factory calibration must only be performed using a special calibra- tion key (Figures 3 and 5G).
Page 6
Be sure to change tips between volume setting adjustments and to pre-rinse each new tip. When the PiPetman Fixed is read- justed to compensate for a particular solution, the original setting (ex. 0.2) can be noted on the self-adhesive labels supplied (Figure 4).
PiPetman diamOnd tips are made to the highest specifications, strict quality control is maintained throughout the manufacturing process. these tips are used to calibrate PiPetman Fixed, therefore for optimum performance we recommend using PiPetman diamOnd tips with your PiPetman Fixed.
Page 8
Filter tips are used when sample-to-sample, sample- to-pipette, or sample-to-operator contamination must be avoided. Gilson's s terilized D iamond F ilter T ips a re c ertified f ree of detectable rnase, dnase, dna, rna, and protease. pIpEtman dIamond tips to use for best results...
PiPetman Fixed. • Make sure to operate the PiPetman Fixed slowly and smoothly. • When aspirating, keep the tip at a constant depth below the surface of the liquid. • The PiPetman Fixed should be held in the vertical position. • Change the tip before aspirating a different liquid, sample or reagent. • Change the tip if a droplet remains on the end of the tip from the previous pipetting operation.
• Dispense the sample into its original vessel. • Ensure that the tip is properly immersed in the sample liquid. • Aspirate the sample more slowly. if the bubble appears a second time, replace the tip. the outside of the PiPetman Fixed can be cleaned using a cloth dampened with isopropanol.
Gilson representative. Before returning a pipette, please ensure that it is completely free of any chemical, biological, or radioactive contamination. 9 - SparE partS Contact your Gilson representa- tive to order spare parts.
Page 13
S.a.S. 19, avenue des Entrepreneurs BP 145 - 95400 VILLIERS LE BEL FranCe Tel.: +33(0)1 34 29 50 00 Fax: +33(0)1 34 29 50 20 Hereby certifies on its sole responsibility that the products listed below: pIpEtman® fixed F2, F5, F10, F20, F25, F50, F100, F200, F250, F300, F400, F500, F1000, F5000 comply with the requirements of the following...