Download Print this page

Gilson SS-18 Operating Manual

Durability index apparatus

Advertisement

Quick Links

ASTM D 3744; AASHTO T 210; CALIF. 227, 229
SAFETY INSTRUCTIONS:
Whether you are the owner, employer, operator, or maintenance
person for this machine, safety is your responsibility. You are
responsible for operating and maintaining this equipment
in compliance with these instructions. You should always
review and completely understand these safety and operating
instructions before using this machine.
Installation and use of the SS-18 Durability Index Apparatus
must conform to all OSHA requirements, the national electrical
code (NEC), and all local codes.
WARNING!
1. DANGER! This machine operates on electrical
current. Improper use could result in electric shock,
electrocution, injury by explosion, or even death.
2. Voltage and moving parts around motors and motor-
driven equipment can cause serious or fatal injuries.
ALWAYS disconnect power source before working
on motor or its connected load.
3. Motor MUST be securely and adequately grounded by
a method approved by the NEC and by local codes.
4. DO NOT use this equipment in a hazardous location
as defined by Article 500 of the National Electrical
Code (NEC).
INSTALLATION:
1. Carefully unpack and inspect the unit for shipping damage.
2. Bolt the SS-18 to the floor or to a sturdy, rigid table.
3. Connect the power cord to a properly grounded power
supply, rated for correct voltage and amperage output.
4. Store unit in a clean, dry, indoor location when not in use.
PHONE: 800-444-1508
740-548-7298
Durability Index Apparatus
SS-18 and SS-18F
P.O. Box 200, Lewis Center, Ohio 43035-0200
E-mail: customerservice@gilsonco.com
OPERATING MANUAL
SS-18 shown with SSA-22
MAINTENANCE:
Read and completely understand all safety instructions before
performing maintenance on the SS-18.
Disconnect power and allow motor to come to a complete stop
before performing maintenance or inspection.
A. Periodically inspect the apparatus, checking for the
following:
1. Dirt accumulation.
2. Unusual noises or vibration.
Website:
www.globalgilson.com
(Continued on back.)
Rev: 09/25/2018
FAX:
800-255-5314
740-548-5314

Advertisement

loading
Need help?

Need help?

Do you have a question about the SS-18 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Gilson SS-18

  • Page 1 1. Carefully unpack and inspect the unit for shipping damage. A. Periodically inspect the apparatus, checking for the following: 2. Bolt the SS-18 to the floor or to a sturdy, rigid table. 1. Dirt accumulation. 3. Connect the power cord to a properly grounded power supply, rated for correct voltage and amperage output.
  • Page 2 6. Overheated electrical connections. is not a hazard. 7. Loose mounting bolts. ADDITIONAL ACCESSORIES AND REPAIR PARTS: 8. If any of these conditions are present, repair the SS-18 before operation. • SSA-22 Washing Vessel • TSA-169R, TSA-169RF Digital Lab Timer...

This manual is also suitable for:

Ss-18f