ENGLISH ORTHOPANTOMOGRAPH™ OP200 D ™ ORTHOCEPH OC200 D Installation & Adjustments Manual 5139799-100 rev. 9 Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 2
Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 3
Instrumentarium Dental. The original language of this manual is English. Instrumentarium Dental reserves the right to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation.
Page 4
Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
software These manuals and future updates are available on request from Instrumentarium Dental. In order to maintain safe and correct operation of the OP200 D unit, read and follow carefully instructions provided in this manual during the unit installation process.
Page 10
WARNING! SERVICING AND INSTALLING THIS UNIT WITHOUT ADEQUATE EXPERTISE IS EXTREMELY DANGEROUS. Instrumentarium Dental recommends that all service operations are performed by Instrumentarium Dental authorized service personnel! WARNING! Prevent unauthorized access to unit during the unit installation process. Some internal parts of the unit contain dangerous voltage, which can be reached during the installation, when the unit covers have been removed.
Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 12
CD-RW Compact disc (read and write) DVD-RW Digital Versatile Disc (read and write) High density drive Peripheral Component Interconnect Local Area Network 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
CERTIFIED TO CSA CERTIFIED TO CSA STD C22.2 NO 601.1. STD C22.2 NO 601.1. 3174420 3174420 Manufactured by Instrumentarium Dental Manufactured by Instrumentarium Dental Nahkelantie 160 Nahkelantie 160 FI-04300 TUUSULA, Finland FI-04300 TUUSULA, Finland Fig 1.2. Main label (230 V) Fig 1.2.
Page 14
1 = OP200 models starting from s/n 100 000 indication of a "Special" version, marked only in products which have a non-standard configuration For example, OP200 D-1-4-1 is: (OP200 D) ORTHOPANTOMOGRAPH OP200 D (-1) with D-051S tube (-4) Removable combined panoramic and cephalostat sensor (-1) first version of OP200.
WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions and maintenance schedules are observed. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
NOTE! Before switching the unit on ensure that the emergency stop switches are released. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
1 Introduction 1.7 Unit identification labels Cephalostat head label Main Label Unit serial number and type label Midsagittal and FH-light laser label 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 19
X-ray generator and tubehead label NOTE! Some labels listed above are related to some options and may not be included in all countries. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Instrumentarium Dental can guarantee OP200 D PCI board and CLINIVIEW software compatibility with certain PC hardware and software only if that configuration has been tested by Instrumentarium Dental. Any later changes to the hardware or software may void this test. 5139799-100 rev 9...
OP200 D has at least the following parts that should be regarded as non-environmental friendly waste products: X-ray source assembly All electronic circuits Column counter weight (Pb). 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
2 Installation requirements and dimensions WARNING! ORTHOPANTOMOGRAPH OP200 D must not be installed and used in rooms where an explosion hazard exists. Equipment not suitable for use in the presence of flammable mixtures. CAUTION! ORTHOPANTOMOGRAPH OP200 D and ORTHOCEPH C200 D units are designed for fixed installation.
NOTE! Install the wall mount bracket in addition to the free standing base plate options to reach the maximum stability. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Controls on the right side of unit. Right side operator’s view for panoramic patient positioning, cephalostat arm on the left. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
OP200 D should have a minimum clearance of 203 cm (80 inches wide) and 155 cm (61 inches) for comfortable patient entry, patient positioning and for servicing the unit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Manual for details. 2.5.10 Mains connection Verify that correct mains voltage is available (3 wire: single-phase, 110 VAC or 230 VAC with protective ground). 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Fluorescent tool OP200 D no. 68662 Torque wrench NOTE! Make sure that all needed tools and items are available before the installation is started! 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 30
2 Installation requirements and dimensions 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Used support material should be high enough to prevent the mains cable to be damaged between the column and floor. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 32
Fig 3.12. Wall mount Fig 3.13. Trim panels Fig 3.14. Mains cable Fig 3.15. Wall bracket and screws mounting screws bracket 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Black colour means that the carton has been dropped and the sensor may have been damaged. If the status shock sensors black, contact manufacturer. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 34
Remove the staples holding the card board bottom flaps. Unfold the flaps. Fig 3.21. Remove styrofoam transport support material located on top of the carriage. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
480 mm (18.9”) on the panoramic side and 1460 mm (57.4”) on the cephalostat side measured from the wall bracket center point. Fig 4.23. Wall mount bracket 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 36
Drill suitable holes according to mounting hardware. Insert plugs if necessary. Fig 4.25. The distance L between wall and column can be adjusted within the limits 160-210 mm. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 37
(part no. 9955) or other strong support on the other side of the wall. The support has to be fixed with threaded fasteners. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Column bottom plate for fork stand installation. Not suitable for floor mounting. Shipped as a part of optional fork stand delivery. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 39
NOTE! When leveled, adjust all remaining adjustment screws against the floor to reach maximum stability. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 41
Attach the anchor plug to the floor. Insert and secure the column with a M6 x 45 mm bolt. Fig 4.45. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 43
Fig 4.50. Fig 4.51. NOTE! Do not fasten the carriage top and bottom covers until instructed. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 44
(X)! Damage to unit and/or degradation of image quality may occur! NOTE! The carriage is heavy. Lifting assistance may be needed. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 45
Insert 2 pcs of M8 x 25 bolts with washers through the carriage fastening holes located near the column pins and tighten the bolts firmly. Fig 4.58. Fig 4.59. Fig 4.57. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
(or by adjusting the base plate, if used). Retighten the bolt. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
CAUTION! Do not try to turn the rotating unit clockwise, as this will twist the cables and will cause damage to the rotating unit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
NOTE! Certain parts of the equipment are covered with protective film to prevent scratches during transportation and installation. Remove all protective films before handing over the unit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 53
CCD sensor as it is located in the same packing as the cephalostat head. NOTE! Do not drop the CCD sensor while unpacking the cephalostat head. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Fig 4.86. Fig 4.87. Fig 4.88. Fig 4.89. Fig 4.90. cephalostat head ready for Fig 4.91. Cephalo end cast cephalostat arm installation 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 55
Use 4 pcs M6 x 16 mm allen key screws (DIN7991). Tighten the screws. Fig 4.95. NOTE! Max. tightening torque 10 Nm/7 ft. lbs. Use torque wrench! 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 56
Fig 4.99. Fig 4.100. Check that cephalostat arm adjustment screw is located on the correct side depending on the unit cephalostat side. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 57
Route the cephalostat cables to the carriage left side cabin. Tighten the cable strain relief. Cable strain relief Fig 4.104. Fig 4.105. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Close the carriage left side cover and fasten the screws. Before closing the cover verify that the cables inside the box are properly routed. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Remove the CCD sensor and secondary collimator transport supports below the cephalostat head. Store them for possible later use. Fig 4.113. Fig 4.114. Fig 4.115. Fig 4.116. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 60
Tighten from above the 2 screws holding the sensor holder to the cephalostat head. Fig 4.118. Tighten also the 2 screws holding the secondary collimator. Fig 4.119. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 61
Loosen the fastening of the cephalostat end cast before the adjustment. Tighten all screws after adjustment is done. Fig 4.123. Fig 4.124. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
OP200 unit manufacturer. More details are found in the OP200 D User Manual chapter OP200 D technical specifications. The chapter 2. Installation 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 63
BLUE BROWN Fig 4.126. NOTE! If changing the voltage selection is not needed, continue at chapter Removing counter weight transport security bolt. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
There is a jumper for line voltage selection available. Remove the jumper from the 230 V position and place it to the 110 V position. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 65
2 and the green wire to pin 3 (ground). Fig 4.131. Power Cord 230 V and Power Cord 110 V. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Store this bolt and the locking plate for future use. Fig 4.134. Counter weight security bolt removal. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Remove the hatch at the lower rear of the column by undoing two M4 screws. Locate the terminal strip connector X103. Fig 4.138. Fig 4.139. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
With your other hand, handle the PCI board by the metal bracket or edges and remove it from the antistatic bag. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Push the sensor handle so that the sensor is locked. Fig 4.142. Placement of the panoramic CCD sensor 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Verify that no error messages appear on the control panel. Connect both optical fibre connectors to the optical fibre terminals in the OP200 D unit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 71
Fig 4.147. Locate the receivers and optical fibre connectors on the back transmitters of the computer. Fig. 4.148.Connect optical fibre connectors to the PCI board terminals 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 72
CCD sensor and PC is working. CLINIVIEW - Help - System Information. Check that the emergency stop switches are working properly. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 73
Fig 4.150. Communication link between CCD sensor and PC is working. Fig 4.151. Communication link between CCD sensor and PC is not working. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Select the VT collimator by using the left / right keys in the control panel. Imaging programs P11 and P12 are lit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 77
Create a test patient. Press “Start image capturing” button to make CLINIVIEW ready for the image capture. Make an exposure. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Remove the tube head covers. Fig 5.160. Fig 5.159. Automatic Automatic collimator and VT collimator aperture. panoramic aperture. Fig 5.161. Fig 5.162. Fig 5.163. Fig 5.164. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 79
Loosen the four screws (B) holding the upper plate (fig 5.168) by ¼ turn and make adjustments using the horizontal adjustment screw (C). 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 80
Use service program Sr 80 CrL. Select the autocollimator channel (kV display: "Co"), one acceleration (mA display: 1), and speed 125% of nominal (s display: 1.25). 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
tolerance of 0.2 mm) The distance between the pins is 60 mm (within a tolerance of +/-1 mm). Fig 5.172. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 82
Tolerance is ±0.2mm. Remove the chin rest, loosen the sinus rest screws and move the sinus rest sideways. Tighten the screws. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
(CLASS 1 LASER PRODUCT). Verify the alignment of the positioning laser lights. 1 = Midsagittal light 2 = FH-light 3 = Layer light 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 84
5 Digital unit adjustments and calibration Fig 5.178. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 85
TMJ pointer, adjust the middle point of the head support knob to the same plane. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 86
Remove the main support cover. Loosen the bolts securing the head support bracket. Adjust the head support bracket and tighten the bolts firmly. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 87
Check also that the lever for Fig 5.191. FH laser light height adjustment moves freely between the lower and upper mechanical stops. Fig 5.193. Fig 5.192. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 88
If the beam is not parallel to the chin rest side groove, adjust the image laser light holder located inside the lower shelf. Fig 5.199. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Make sure the AEC calibration tool is in front of the panoramic collimator and that the front tube head cover is in place while measuring. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 90
5 Digital unit adjustments and calibration After the AEC frequency is adjusted, re-install the rotating carriage bottom cover, tube head covers and the head rest. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Tighten the locking screw after the adjustment is done. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
- Make a new exposure and repeat this step as many Fig 5.206. Correct rotational alignment. Note the line drawn as times as needed. reference. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 93
- Tighten the locking screw (LOCK in Figure 7) when you are done adjusting. Fig 5.209. Correct vertical alignment. Note that the imge’s contrast is set to its maximum value. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 94
Fig 5.212. Enter distance to left border in the kV field and distance to right border in the mA field. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Tighten the screws. Verify by taking a test exposure. Fig 5.214. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 96
Avoid moving the gear along its axis when adjusting the secondary collimator. Fig 5.218. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 97
NOTE! Do not adjust anything at the tube nor change the AL setting! Only the cephalostat secondary collimator may be moved around! 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
H1 nor H3 should illuminate except shortly after exposure button is pressed. Check above procedure also with imaging program P9b. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 99
During exposure these leds mirror the status of leds H1, H2 and H3 on the beam alignment board. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
CCD sensor side. If the pins cross each other, the ear holders must be adjusted. Fig 5.223. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 101
When the horizontal adjustment is complete, tighten all locking screws and re-fasten the string. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 102
When the horizontal adjustment is complete, tighten all locking screws and re-fasten the string. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Adjust with the trimmer R146 (d) the reading of the frequency meter to 2000 Hz (± 25 Hz). Check the readings again and repeat the adjustment if needed. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Place the blue cover against the white one. Lift upwards and tighten screw bottom secondary collimator. Fig 5.239. patient positioning panel connector 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 105
Fig 5.243. Fig 5.244. Install the cephalostat head top cover and tighten the screws located in all four corners. Fig 5.245. Fig 5.246. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 106
5 Digital unit adjustments and calibration 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Fill out the documents required. For future purposes, it is advised to fill out the OP200 D Configuration Form. Dispose of packages and clean up the installation site. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
The kV and mA values are not shown. 6.3.4 QA-image The Orthopantomograph OP200 D can produce a quality assurance image for checking that OP200 D digital system (OP200 D unit, PC hardware with connections and CLINIVIEW software) works properly.
Page 110
If everything is as described, the system is working properly. Set the primary collimator back to the panoramic position and insert the bite block. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Place Test Tool Digital (part no. 68772) and ball pin phantom (part no. 68799 - supplied with the main unit) at the chin support. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 112
NOTE! Line pair resolution measured with this QA test is not directly comparable to line pair resolution measured with clinical patient images. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 113
Fig 6.252. Digital Test Tool and test tool holder Launch CLINIVIEW software, select (or create) an appropriate "TEST IMAGE" patient and enable image capturing. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
See OP200 D Maintenance Manual. Briefly review user manuals. Leave this OP200 D Installation & Adjustments Manual and OP200 D User Manual at the site. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
CLINIVIEW software settings to go with each PC monitor’s characteristics. Details on modality specific settings are found in the CLINIVIEW User Manual. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 116
6 Handing over the unit 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Clean the unit. Give user training. Check several examples of the customers images to verify correct operation of the unit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 118
7 Appendix A, OP200 D installation check list 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Remove Frankfurt laser light by opening three screws. Fig 8.253. Loose the FH laser light cable from LRD board connector. Fig 8.254. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 120
Fig 8.256. Fig 8.257. Re-route the FH laser cable under the CPU board right lower corner behind the board fastening screw. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 121
Fig 8.259. Re-route the FH laser cable out of carriage middle cabin and connect back to the I/O board connector. Fig 8.260. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 122
Attach plastic blugs to empty side. Fig 8.264. Fig 8.265. Fig 8.266. Fig 8.263. Cabin door Control panel Plastic blugs Plastic blugs screws for control panel holder 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Fig 8.267. Cephalostat end cast Fig 8.269. Cable fastening plate Fig 8.270. Cable with the connector Fig 8.271. Right handed cephalostat head and cable routing 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 124
Fig 8.273. Column end cast Fig 8.274. Right handed cephalostat arm adjustment and re-inforcement plate Fig 8.275. Left handed cephalostat arm adjustment 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 125
Fig 8.278. I/O board and location of X15 jumper. Adjust the cephalostat adjustments according to the chapter 5.3. Optional cephalostat adjustment and calibration. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 126
8 Appendix B, Unit orientation change NOTE! Before adjustments check that cephalostat arm and head are levelled. See chapter 4 Hardware installation. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
NOTE! The content of the cephalostat upgrade kit may vary depending on unit version and options. NOTE! Check the packing list delivered with the upgrade kit. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Fig 9.281. Switch off the unit power. Remove the counter weight hatch cover behind the column. Fig 9.282. Fig 9.283. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 130
Counter weights OP200 D 4 small weights 1 large weight OC200 D 4 small weights 24 large weights Steel plate with plastic guides 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Page 131
Gap between carriage and column should be equal on both sides of the unit. Fig 9.287. Gap between carriage and 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
5.164). Set the old tubehead cover aside. Move the collimator stopper plate to the side so that it is possible to select cephalostat slot. Fig 9.289. Cephalostat slot 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
First loosen the sensor and secondary collimator screws and install the supporting brackets. Fig 10.292. Transport supporting brackets 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
See the figures below for instructions how to re-pack the internal packaging parts of the cephalostat package. Fig 10.302. Fig 10.302. Fig 10.302. Fig 10.302. Fig 10.303. Fig 10.303. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...
Remove the panoramic CCD sensor and place it steadily on the corresponding packaging part. Fig 10.304. Fig 10.304. Fig 10.304. Fig 10.304. 5139799-100 rev 9 Instrumentarium Dental Reviewed Honkala Jorma Tapio 2018-05-22 14:56 D500316, 9 Approved Ukkonen Juha-Pekka 2018-05-23 14:32...