Sitting Positioning In Bed; Cpr (Cardiopulmonary Resuscitation); Transport Function; Pack&Go® Function - Care of Sweden CuroCell M4 Instructions For Use Manual

Table of Contents

Advertisement

Note!
• Make sure that the mattress system is
filled, which is shown by a green light
from the diode, before performing a
hand check.
• How to perform a hand check depends
on which mattress is used.
When using an overlay mattress system
(CuroCell® OP10 or CuroCell® CX10)
1a. Open the cover and identify a cell with less
air at the user's sacrum. Insert a hand, with the
palm facing up, between the overlay mattress
and the underlying mattress. The hand is
inserted beneath the user's sacrum (center of
mattress).
When using a full replacement mattress
(CuroCell® CX15, CuroCell® CX16 or CuroCell®
CX20)
1b. Open the cover and identify a cell with less
air at the user's sacrum. Insert a hand, with the
palm facing up, between the top cells and the
underlying safety mattress. The hand is
inserted beneath the user's sacrum (center of
mattress).
2. Ensure there is a gap between the user and
the underlying mattress so that the user does
not 'bottom out'.
3. If you can feel the user's sacrum resting in the
palm of your hand, the gap is too small. See
section 8 'Troubleshooting'.
5.7

Sitting positioning in bed

See 3.1

5.8 CPR (Cardiopulmonary resuscitation)

Se 3.2
5.9

Transport function

Se 3.3
5.10
Pack&Go® function
The product may be packed as follows:
Ensure that no-one is lying on the
mattress.
On the control panel, press the
lock/unlock button.
Press the Pack&Go® button and hold it
down for 2 seconds
The Pack&Go® diode will flash during deflation.
The mattress will empty of air and be ready to
be simply folded together within 20 minutes.
The control unit gives an audio signal once
deflation is complete.
Carefully fold the mattress together, place the
control unit between the folds of the mattress
and place the system in a transportation bag
(accessory) or equivalent for protective storage.
Ensure that the power supply is packed
complete.
5.11

Restart

See 3.4
5.12

Power failure

See 3.6
5.13

Notifications

Different notifications exist based on
how serious the warning is. With a
malfunction or an error, a notification
will be given by way of a flashing warning
triangle. To mute the warning signal, press the
mute button.
and a notification code is shown on the four
different cycle time diodes (10, 15, 20, 25). To
read the cycle time during the error notification,
unlock the control panel.
5.14

Maximum pressure notifications

See 3.5
5.15

Table of notifications

Information about each notification is shown in
the notification table:
Notifications from 1-10 are both audible and
visual. The error code will be displayed until the
error has been rectified. If the mute button is
pressed the audible warning will crease for a
period of 5 minutes and will return until the
error has been rectified. The notifications 11-13
have no audible alarm. The error code is shown
until the system is restarted.
14
CuroCell
When a notification
occurs, the current cycle
time diode will turn off
®
A4