Hand Check (Function Control); Sitting Positioning In Bed; Cpr (Cardiopulmonary Resuscitation); Transport Function - Care of Sweden CuroCell M4 Instructions For Use Manual

Table of Contents

Advertisement

4.6

Hand check (function control)

Hand check is performed to ensure that the
mattress system works properly and to ensure
that the weight setting is correct. Hand check
should be performed regularly; for CuroCell®
M4, we recommend once per work shift as well
as after installation of the system.
For manual systems hand check should also be
performed when moving from lying to seating
position in bed, in case of positioning changes
and when changing the comfort setting (and/or
the weight setting).
Note!
• Make sure that the mattress is filled, and
that the system has adapted to the user
before performing a hand check.
• How to perform a hand check depends
on which mattress is used.
When using an overlay mattress system
(CuroCell® OP10 or CuroCell® CX10):
1a. Open the cover and identify a cell with less
air at the user's sacrum. Insert a hand, with the
palm facing up, between the overlay mattress
and the underlying mattress. The hand is
inserted beneath the user's sacrum (center of
mattress).
When using a full replacement mattress
(CuroCell® CX15, CuroCell® CX16 or CuroCell®
CX20)
1b. Open the cover and identify a cell with less
air at the user's sacrum. Insert a hand, with the
palm facing up, between the top cells and the
underlying safety mattress. The hand is
inserted beneath the user's sacrum (center of
mattress).
2. Ensure there is a gap between the user and
the underlying mattress so that the user does
not 'bottom out'.
3. If you can feel the user's sacrum resting in the
palm of your hand, the gap is too small. See
section 8 'Troubleshooting'.
4.7

Sitting positioning in bed

See 3.1
4.8

CPR (Cardiopulmonary resuscitation)

See 3.2
4.9

Transport function

See 3.3
4.10
Pack&Go®
The product may be packed as follows:
Ensure that no-one is lying on the
mattress.
On the control panel, press the lock/unlock
button.
Press the Pack&Go® button and hold it
down for 2 seconds
The Pack&Go® diode will flash during deflation.
The control unit gives an audio signal once
deflation is complete.
Carefully fold the mattress together, place the
control unit between the folds of the mattress
and place the system in a transportation bag
(accessory) or equivalent for protective storage.
Ensure that the power supply is packed
complete.

4.11 Restart

See 3.5

4.12 Power failure

See 3.6
4.13

Notifications

Different notifications exist based on
how serious the warning is. With a
malfunction or an error, a notification
will be given by way of a flashing
warning triangle. To mute the warning
signal, press the mute button.
will turn off and a notification code is shown on
the four different cycle time diodes (10, 15, 20,
25). To read the cycle time during the error
notification, unlock the control panel.

4.14 Maximum pressure notification

See 3.5
9
CuroCell
When a notification occurs,
the current cycle time diode
®
M4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Curocell a4Curocell iq

Table of Contents