Nanoleak Operation - Redline NANOLEAK FINDER AIR BRAKE Operation Manual

Table of Contents

Advertisement

NANOLEAK
LEAK SIZE INDICATOR
The digital leak size indicator remains off normally but once a leak is detected, a number from 1-9 will be displayed
for all NanoLeak Signature Air regardless of the sensitivity setting.
The number will continue to increase or decrease depending on the amount of NanoLeak Signature Air that is
detected. The maximum value will be displayed once the leak source has been located. The table below can be
used to approximate the size of leak :
Maximum # Displayed Approximate Leak Size (oz/vr)
LOW BATTERY INDICATOR
Replace the 4 AA alkaline batteries when the red LED on, the control panel is lit. Follow battery installation
instructions under Maintenance section.
AUDIO MUTE FUNCTION
To silence or mute the audio beep and alarm signal, press the MUTE button. To restore the audio sound, press
the MUTE button again. (Note: a few seconds is required to restore sound if the mute button is pressed in rapid
succession.)
ADJUSTING SENSITIVITY LEVELS
The NanoLeak
Handheld Leak Locator will default to the NORMAL sensitivity level automatically once the unit
comes out of the warm up cycle and the green LED will turn on.
To change sensitivity levels, press the SENSOR once for HIGH sensitivity (red LED will turn on) and again for LOW
sensitivity (yellow LED will turn on).
LEAK TEST VIAL
The NanoLeak
Handheld Leak Locator comes with a Leak Test Vial that allows the user to make sure the
NanoLeak
Handheld Leak Locator is performing properly. To test:
1. Remove the plastic seal cap on top of the Leak Test Vial by pulling it off (see fig. below).
2. Turn on the locator and allow the unit to complete the warm up cycle .
3. Place the sensor close to the small hole in the top of the Leak Test Vial. The beep rate should increase and
the Digital Leak Size Indicator should display a number from 4-6 indicating that the sensor and electronics are
working properly.
10
HANDHELD LOCATOR OPERATION
Maximum # Displayed
1-3
4-6
7-9
Remove Seal Cap To test
Leak Test Vial
Approximate Leak Size (oz/vr)
< 0.1
0.1 to 0.5
> 0.5
NOTE: Always remember to replace
plastic seal cap after leak test is
completed. Replace Test Vial when
the green color is no longer visible.