Control Panel And Keypad - Sanyo MDF-U333 Instruction Manual

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

FREEZER COMPONENTS

Control panel and keypad

< MDF-U537D >
1. Alarm test key (ALARM TEST): Test key for alarm device. By pressing this key, the alarm lamp is
flashed, remote alarm is activated and buzzer sound. This means all alarm function operate correctly.
2. Buzzer stop key (BUZZER): To silence the audible alarm, press this key. The remote alarm is also
silenced by pressing this key. (The buzzer cannot be stopped during remote alarm testing.)
3. Defrost key (DEF):
Pressing this key again after defrosting leads resumption of the operation.
Note: The refrigerating operation never resume automatically after defrosting.
4. Set key (SET): Temperature setting mode is led by pressing this key. Once the key is pressed, the
changeable digit is flashed. Pressing this key again after setting desired temperature, the setting is
stored into computer memory. If there is no key operation for 90 seconds during the temperature setting
mode, the temperature setting mode is invalid automatically. See page 19 for the details.
5. Digit shift key (
lock is available by pressing this key for more than 5 seconds in the temperature display mode. Refer to
page 20 for the key lock.
6. Numerical value shift key (
shift. "ON-OFF" of key lock can be selected by pressing this key in the key lock mode.
7. Alarm lamp (ALARM): This lamp is flashed when the audible alarm is activated.
8. Digital temperature indicator:
temperature.
9. Compartment select key (TOP/BOTTOM) (MDF-U537D only):
temperature display of top compartment or bottom compartment and each setting is selected.
10. Top compartment lamp (TOP) (MDF-U537D only): This lamp lights when the top compartment is
selected.
11. Bottom compartment lamp (BOTTOM) (MDF-U537D only): This lamp lights when the bottom
compartment is selected.
7
1
2
3
By pressing this key for 5 seconds, the refrigerating operation is stopped.
): Pressing this key in the setting mode causes the changeable digit to shift. Key
): Pressing this key in the setting mode causes the numerical value to
This indicator shows the present chamber temperature or set
11
10
4
9
10
8
5
6
By pressing this key, internal

Hide quick links: