Sanyo MDF-U333 Instruction Manual page 18

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

OPTIONAL COMPONENTS
Model MDF-U537D:
< Attachment of MTR-G85 on the left lower front of the freezer >
1. Attach the recorder to the mounting kit by following
the procedure shown in the manual enclosed with the
mounting kit.
2. Remove the screw on the cover for the recorder
space on the lower front of the freezer. Open the
cover and push the shaft on the both sides outward to
remove the cover. Fig. 14
3. Install the recorder in the recorder space.
4. Join the wire in the unit compartment with the
recorder wire by each wire connector. See Fig. 15
4. Remove the wire grille on the back bottom of the
freezer. See Fig. 16
5. Route the capillary tube of the recorder to the back
of the freezer through the unit compartment
6. Remove 2 rubber caps (outside and inside) on the
access port on the back of the freezer and also
remove the insulation in the port.
7. Pass the recorder sensor into the chamber through
the access port. Fig. 17
8. After covering the recorder sensor with the cover
provided, fix the recorder sensor under the middle
shelf by using 2 enclosed binders. See Fig. 18.
9. Make a small cut on the rubber caps for capillary to
pass.
10. Replace the insulation into the access port and
cover the port completely with the rubber caps.
Access port
Fig. 17
17
Shaft
Fig. 14
Fig. 15
Fig. 16
Sensor
Fig. 18

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-u537Mdf-u537d