Remote Alarm Terminal - Sanyo MDF-U333 Instruction Manual

Biomedical freezer
Hide thumbs Also See for MDF-U333:
Table of Contents

Advertisement

REMOTE ALARM TERMINAL

WARNING
Always disconnect the power supply cord before connecting an alarm device to the remote alarm
terminal.
The terminal of the remote alarm is installed at the back of the unit. The alarm is generated from this
terminal. Contact capacity is DC 30V, 2 A.
Contact output:
At normal condition "Open"
At abnormal condition "Close
Note:
The remote alarm is silenced by pressing the BUZZER key as the remote alarm is operated in
conjunction with the buzzer except for the power failure alarm and alarm test.