3-2.Air Velocity Analog Output Setting
The output can be selected from the voltage output (0 to 5 V) and the current
output (4 to 20 mA).
The range of air velocity is also selectable.
The factory setting is: DIP Switch setting 4; air velocity (0 to 50.0 m/s); and current output.
DIP switch
setting
0
1
2
3
4 to 9
Example: For the probe 0975-31, air velocity 10.0 m/s, DIP Switch 3 (Range: 25.0 m/s)
While changing the output setting, be sure to shut off power supply.
20
Velocity range
Voltage output range
[m/s]
0 to 2.0
0 to 5.0
0 to 10.0
0 to 25.0
0 to 50.0
Current output range
[V]
0 to 5
0 to 5
0 to 5
0 to 5
0 to 5
[mA]
4 to 20
4 to 20
4 to 20
4 to 20
4 to 20