Test Procedures; Test Setup - Cal Test Electronics CT4447 Series User Manual

30 mhz high-voltage differential probes, ctuvus listed
Table of Contents

Advertisement

CT4447 Series High-Voltage Differential Probes User Manual

Test Procedures

Test Setup

1.
Turn on the oscilloscope.
2.
Connect the probe to any channel on the oscilloscope (for warm-up).
NOTE
It is recommended to operate the probe using USB power during verification testing.
Il est recommandé d'utiliser la sonde en utilisant l'alimentation USB pendant les tests
REMARQUE
de vérification.
3.
Connect the USB cable's to the probe's USB-C power input jack and the AC power adapter to to
the cable's USB-A connector. Plug the AC adapter into mains line voltage.
4.
Turn on the probe and verify the LED lights.
5.
Turn on the remaining test equipment and let the probe and equipment warm up for a minimum of
20 minutes.
23