After confirming that the probe operator is not touching the device under test, turn on the high-volt-
age source.
9.
Measure the voltage under test and observe the waveform on the oscilloscope.
WARNING
Remember the actual voltage is the attenuation factor greater than the oscilloscope
waveform.
N'oubliez pas que la tension réelle est le facteur d'atténuation supérieur à la forme
AVERTISSEMENT
d'onde de l'oscilloscope.
10. Turn off high voltage source.
11. Disconnect the probe inputs from the high-voltage source.
CT4447 Series High-Voltage Differential Probes User Manual