Summary of Contents for Bante Instruments SALscan10
Page 1
SALscan10/20 Pocket Salinity Tester USER MANUAL Bante Instruments Inc.
Page 2
SALscan10/20 MEAS Indicates that the tester is in the measurement mode Overview Indicates that the tester is in the calibration mode Thank you for selecting the SALscan series pocket salinity tester. This SETUP Indicates that the tester is in the setup mode...
Page 3
SALscan10/20 Setup Menu Option Menu item The SALscan series tester contains 6 menu items in the setup menu, the following table describes the functions of each option. Menu Item Option and Description If necessary, press the Cal key again to select an option.
Page 4
SALscan10/20 SALscan20 Measurement Range Standard Solution Range Default 200 to 2000 µS/cm 700 to 1700 µS/cm 1413 µS/cm 2 to 20 mS/cm 7 to 17 mS/cm 12.88 mS/cm 20 to 200 mS/cm 70 to 170 mS/cm 111.8 mS/cm Make sure that using the fresh standard solution during the calibration.
Page 5
SALscan10/20 Measuring the Sample Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently, make sure that no air bubbles on the sensor surface. Wait for the measurement to stabilize and record the reading.
Page 6
LCD, 21 × 21 mm (0.82 × 0.82 in.) Order Code Description Power Requirements 2 × 1.5V AAA alkaline batteries For SALscan10 tester, cell constant K=1, E-ECscan-C1-10K Auto-Off 8 minutes after last key pressed measuring range 10 to 20 mS/cm Dimensions 185 (L) ×...