General Description - Cole Parmer INC-200 Series Instruction Manual

Table of Contents

Advertisement

Fault
Indicator
Pre-set
Temperature
Figure 1: INC-200 controls

General Description

The INC-200 and INC-200D incubators are
constructed from clear acrylic material for total
visibility. The units are designed for easy access
with a hinged front door panel. Each side panel
has 2 x 10mm diameter plugged holes for the
introduction of either gases or cables. The
incubator does not have a base so that it can be
placed directly over complete instruments (an
optional base and shelving system are available
as accessories).
The interior temperature is controllable up to
60ºC; forced air circulation and electronic tem-
perature control ensure accurately maintained
conditions. Additionally there is a temperature
safety cut-out set at 72ºC.
The INC-200 has an analogue controller while
the INC-200D has a digital display system. In ad-
dition to the main temperature selector there is a
push switch temperature re-set which maintains
the incubator at 37ºC.
Controls
Refer to Figures 1 & 2 to identify the controls.
Mains On/Off: This is a rocker-type switch. Pressing
the top switches the power on and pressing the
bottom switches the power off. The switch is
illuminated when the unit is on.
Temperature
Control
Load
Indicator
Mains
On/Off
Temperature
Figure 2: INC-200D controls
Pre-set Temperature Switch: This is a rocker-type
switch. Pressing the top switches the incubator
to the pre-set temperature (37ºC). Pressing the
bottom switches the incubator to the variable
setting.
Load Indicator: An amber LED illuminates when
the heater is in operation. This light remains on
while the incubator is reaching the set
temperature and flashes when the set
temperature is reached.
Fault Indicator: This is a red warning light. When
lit it indicates a fault has occurred and the
temperature safety cut-off has been initiated (set
at 72ºC).
Set Switch: This is a plain colour rocker switch.
Pressing and holding the switch down displays
the set temperature.
Temperature Control: This knob is used to set the
temperature of the incubator in conjunction with
the press to set switch.
Digital L.C.D.: This is used to display both the
actual and set temperature of the incubator in
ºC. The range is ambient +5ºC - 60ºC with a
resolution of 0.1ºC (INC-200D only)
Set
Switch
Fault
Indicator
Pre-set
Digital
Display
Temperature
Control
Load
Indicator
Mains
On/Off

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inc-200d

Table of Contents