Contents 1 Product introduction 1.1 Preface 1 Product introduction ................1 BA International is a leading brand of dental equipment and handpieces. Our products are produced to high standards and with strict 2 Installation .................... 5 quality controls. To find out more about the rest of our product range, 3 Function and operation of product ............
Page 3
1.5 Intended Use Optima E+ BAE380R 1.5.1 The device can be used for preparation and enlargement of root canals, or device for measuring canal length. 1.5.2 The device must be operated only in hospitals and clinics by qualified dental professionals. 1.6 Contraindication a) Doctors with a pacemaker are forbidden to use this device.
1.7.10 Do not remove the contra angle before the motor handpiece 1.9.3 Torque range: 0.4Ncm-5.0Ncm ( 4mNm ~ 50mNm ) stops rotating. Otherwise the contra angle and the gear inside motor 1.9.4 Speed range: 100rpm~2500rpm handpiece will be broken. 1.9.5 Wireless charging 1.7.11 Please confirm that the file is properly installed and locked in Frequency range: 112-205KHz place before starting the motor handpiece.
Page 5
2.2.1.4 SGP Mode Safety Glide Path Mode F: Forward angle, R: Reverse angle Measuring wire (BA182619) File clip (BA182621) The rotation angle is adjustable, but the forward angle must be equal to the reverse angle. 2.2.1.5 ATR Mode ATR: Adaptive Torque Reverse function. Lip hook (BA182620) Touch probe (BA182622) Disposable insulation sleeves (BA182623)
Page 6
positioning slot locating pin The meter numbers 1.0, 2.0, 3.0 and digital numbers 00-16 do not represent the actual length from the apical foramen. It simply indicates the file progression towards the apex. The digital numbers -1 and -2 indicate that the file has passed the apex foramen. The digital number “00” indicate that the file has reached the apex foramen.
Page 7
Turn the file back and forth until it is lined up with interior latch groove 2.6 Canal measurement functional connection and slips into place. Release the button to lock the file into the contra This is not required if the canal measurement function will not be angle.
2.7 Installation and removal of disposable isolation sleeves e. Preset program selection Long press Setting button “P” to enter preset program during standby 2.7.1 Installation state, press Adjusting button “+”/“-” to select file system , press Setting Before each use of the handpiece and after the handpiece is cleaned button “P”...
Canal measurement state interface Shows the point inside the canal where specified Flash Bar Position a. Canal length indicator bar apical action is triggered. b. Indication number The file rotation starts automatically when the file Auto Start Digital numbers 00-16 do not is inserted in the canal.
Page 10
c) Press the Main button again, and then the motor handpiece returns The speed setting can be adjusted from 100 rpm to Standby interface. to 2500 rpm. d) Hold down the Setting button “P”, then press Main button to turn Press Adjusting button “+”/“-”...
Page 11
Rotation starts automatically when the file is inserted into the canal and the canal length indicator bar lights up more than 2 bars. Forward Angle .In the SGP mode, the Forward P ress Adjusting button “+”/“-” to change. Angles of 20°~400° are available. OFF: Motor does not start when file is inserted In the ATR mode, the Forward Angle of into the canal.
Page 12
The parameters of the presets can also be Press Setting button “P” again, the “Auto changed make it different from default setting. Standby Scr” can be changed, press Adjusting If you want to change back to default setting, button “+”/“-” to adjust, then press to "Main" long press Setting button “P”...
Page 13
the file would return to normal rotation mode when the load is below the Motor combined with canal preset torque value again. measurement function mode When using motor combined with canal measurement function, the measuring wire must be connected with motor handpiece by USB socket, and white socket connects with patient’s lip by lip hook, keep the black socket idle.
Page 14
4.8 Canal measurement operation Connection testing It is strongly recommended to test the When using in standalone apex locator connection every time before use. Clip mode, we suggest putting the motor the holder onto lip hook and check that handpiece on the charging base to get all the bars on the meter on the screen better visual angle.
Page 15
Broken crown. Extremely dry canal. If the crown is broken and a section If the canal is extremely dry, the meter of the gingival tissue intrudes into the may not move until it is quite close to cavity surrounding the canal opening, the apex.
a) Turn the motor handpiece power off. b) Use tweezers etc. to open the rubber cover and then remove the 5 Troubleshooting screw. Failure Possible cause Solutions c) Remove the battery cover. The motor handpiece Motor is on EAL mode, Change to CW, CCW, d) Remove the old battery and disconnect the connector.
Page 17
Do not use chloride detergent materials. a)Use distilled water or deionized water for cleaning at room 6.2.3 Do not use bleach or chloride disinfectant materials. temperature. 6.2.4 For your own safety, please wear personal protective equipment 6.3.1.2 Manual disinfection steps: (gloves, glasses, mask).
Page 18
6.4 The cleaning, disinfection and sterilization of contra-angle,lip Tools: tray, soft brush, clean and dry soft cloth. 1. Remove the shanks/files. hook, file clip,protective silicon cover,touch probeare as follow. 2. Remove the file clip, isolation sleeve,Contra-angle and connecting Unless otherwise stated, they will be hereinafter referred to as wire from the handpiece in sequence, and then put them into a clean tray;...
Page 19
• The cleaning procedure is suitable for the product, and the irrigating carried out together. period is sufficient. c) Cleaning: (c1) The cleaning procedure should be suitable for the It is recommended to use a washer-disinfector in accordance with product to be treated. The flushing period should be sufficient (5-10 EN ISO 15883.
Page 20
6.4.6 Inspection and maintenance 6.4.8 Sterilization 6.4.6.1 Inspection Use only the following steam sterilization procedures (fractional pre- In this chapter, we only check the appearance of the product. vacuum procedure*) for sterilization, and other sterilization procedures 1. Check the product. If there is still visible stain on the product after are prohibited: cleaning/disinfection, the entire cleaning/disinfection process must be ·The steam sterilizer complies with EN13060 or is certified according...
Notes: 7.3.3 Keep out of direct sunlight and avoid getting wet in rain and a) The storage environment should be clean and must be disinfected snow during transportation. regularly; 8 Environmental protection b) Product storage must be batched and marked and recorded. Please dispose according to the local laws.
Technical Description Concerning Electromagnetic Emission Humidity limitation Temperature limitation Table 1: Declaration - electromagnetic emissions Atmospheric Guidance and manufacturer’s declaration - electromagnetic emissions CE marked product pressure for storage The model BAE380R is intended for use in the electromagnetic environment specified below. The customer or the user of the model BAE380R should assure that it is used in such an environment.
Page 23
Voltage dips, short <5 % UT <5 % UT Mains power quality should be NOTE I At 80 MHz end 800 MHz. the higher frequency range applies. interruptions and (>95% dip in UT.) (>95% dip in UT.) that of a typical commercial NOTE 2 These guidelines may not apply in all situations.
Apex Locator Troubleshooting This guide is suitable for the users who use B.A. International’s apex locator for the first time, and those who get unstable readout because of improper operation. Problems Possible Causes Analysis Solutions No readouts File clip File clip wire broken or poor contact Replace the file clip or unstable Measuring wire Measuring wire broken or poor contact...
Page 26
B.A. International Ltd. Unit 9, Kingsthorpe Business Centre Studland Road, Northampton NN2 6NE Tel: +44 1604 777700 Web: www.bainternational.com Guilin Woodpecker Medical Instrument Co., Ltd. Information Industrial Park, Guilin National High-Tech Zone Guilin, Guangxi, 541004 P.R. China MedNet EC-Rep GmbH...