Vesta Charger Wireless Nominal Specifications
Characteristic
MICS MedRadio
Frequency Band
Bandwidth
Modulation
Radiated Power
Range
Transcutaneous Energy Transfer
Frequency Band
Bandwidth
Modulation
Radiated Power
Range
OPTIMIZER Smart Mini IPG Wireless Nominal Specifications
Characteristic
OPTIlink MICS MedRadio
Frequency Band
Bandwidth
Modulation
Radiated Power
Range
402 – 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
< 145 kHz
FSK
< 25 µW E.I.R.P.
0 to at least 1.5 m
13.56 MHz
Industrial, Scientific, and Medical radio
band (ISM)
< 0.014 MHz
Amplitude (slow to optimize coupling, no
data transmitted)
< 0.6 W
5 mm to 40 mm
402 – 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
< 145 kHz
FSK
< 25 µW E.I.R.P.
0 to at least 1.5 m