ROUGHNESS TESTER
OPERATION MANUAL
V0

Advertisement

loading
Need help?

Need help?

Do you have a question about the ISR-S300 and is the answer not in the manual?

Questions and answers

Summary of Contents for insize ISR-S300

  • Page 1 MN ISR www.insize.com ISR-S300 ROUGHNESS TESTER OPERATION MANUAL...
  • Page 2: Product Introduction

    To get the best results from this instrument and to use it General descripition: satisfactorily, first read this manual. Observe the following The ISR-S300 palm-top roughness tester is a portable, light and precautions to obtain the best results from the instrument in the compact instrument, extremely easy to handle and use.
  • Page 3 All the settings are never lost even after the battery has been the main menu, causes the higher level menu to return on display. completely discharged since the instrument uses a particular 4. Scroll key. If pressed when the result of a measurement is on MRAM internal storage memory.
  • Page 4: Menus And Settings

    Switching on the instrument The roughness tester is switched on by pressing the key . The ISR-S400, after a number of internal checks have been conducted quickly, indicates on the display for a few seconds before moving on to displaying the parameters. This indicates to the operator that the instrument is ready to be used.
  • Page 5 The display lists the options possible, while the current selection is indicated by the black background, for example, . To modify it, set the selection on the value desired. The traverse length made by the traverse unit corresponds to the length λ t which depends on the cut-off selected, according to the relation: ( cut-off no.+1)λ...
  • Page 6 The main menu is activated by pressing the key , when the Controlling the stylus position: display shows the cut-off, a parameter, or an error signal. The Make your selection on the item Position and press menu is presented in the following way with the last item selected following indication will be obtained where the vertical with black background.
  • Page 7 Note if you need to check the value of a disabled parameter Setting the calculating standard: and you have just finished measuring simply enable the In the main menu make your selection on the item Standard and parameter concerned. It is not necessary to complete the press The display will show measurement.
  • Page 8 2) Move the selection on the desired item then press Press for interrupting the visualization and to go back to the enable it. The item is enabled when to its left the symbol menu Files is displayed whereas it is disabled when to its left the symbol Eliminating all the measurements stored: is displayed.
  • Page 9 Note: Measurements should always be made with the You are recommended not to use the specimen plate more than is temperature of the room, of the part and of the instrument strictly necessary, to avoid deterioration caused by possible constant and, where possible, at 20°C. scoring of the stylus which could falsify the roughness values.
  • Page 10: Error Messages

    Signalled when the data are required for printing or displaying or when the roughness tester has to receive external commands. The report is generated by the instrument timing-out for 2 seconds. Probable causes: a) The instrument that has to receive the data is not connected or is switched off.