4. The printer will continue with the 'Select for Modality' screen, allowing
you to change the Look-up table, the Interpolation and the maximum
density (Dmax) values manually.
<Modality name>
<Modality name>
<Modality name>
3 Dmax
3 Dmax
3 Dmax
5. In the 'Select for Modality' screen, press the Confirm key to select 'Look-up
table'
The 'Look-up table' screen is displayed:
<Modality name>
1*Linear
<Modality name>
<Modality name>
2 Kanamori
1*Linear
1*Linear
2 Kanamori
2 Kanamori
3 Kanamorilike (100)
3 Kanamorilike (100)
3 Kanamorilike (100)
6. Press the up/down key to select the desired look-up table.
7. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
In case Kanamorilike (100) is selected, you can change the Kanamorilike
value (the range is: 75-220, the default value = 100).
An additional screen for the Kanamorilike value is displayed:
<Modality name>
<Modality name>
<Modality name>
Enter the Kanamorilike value by means of the arrow keys. Press the
Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
Related Links
Controlling the printer via a remote PC (with browser)
Accessing the Change settings menu
Data entry
on page 12
Removing a modality
Adding a modality
Changing Image quality settings for a modality - Changing the
interpolation
1. On the Change settings menu, press the Down key four times, followed by
the Confirm key to select 'Image quality'.
Dry 2.4M | Advanced Operation (key-operator mode) | 41
Note: In case the image quality parameters need to be
changed for this modality, please use the DICOM
Configuration tool in the Key-operator mode via a remote
connection. Alternatively, you can also remove this modality
and add a new modality with the same name. For this new
modality new parameters can be assigned via the local
keypad.
on page 23