Technical specifications of the image plates

Sizes
35 x 43 cm (14 x 17")
35 x 35 cm (14 x 14")
24 x 30 cm
18 x 24 cm
8 x 10"
10 x 12"
15 x 30 cm
Plate construction
Protective layer
Phosphor
Base
Characteristics
Its luminescence spectrum is the typical Eu
around 390 nm in lattices of the BaFBr-type. The top in the luminescence
spectrum is shifted slightly to longer wavelengths due to the incorporation of
iodide.
The stimulation spectrum is much broader than that of pure BaFBr and is
shifted to longer wavelengths. This shift is caused in the first place by the
partial replacement of Ba by Sr, and in the second place by the incorporation
of iodide. Thanks to the red-shift of the stimulation spectrum, maximum
stimulability is assured at 633 nm, the wavelength of the stimulating laser.
The Agfa phosphor has excellent dark decay characteristics. Two hours after
exposure, approximately 80% of the energy stored upon exposure is still
available. The image retention is greater than 50% up to 24 hours after
irradiation.
4451C EN 2013-04-30
Electron beam cured polymer
BaSrFBrI:Eu
P.E.T.
2+
-luminescence, which is at
CR 85-X D
IGITIZER
111
CR cassettes