P/N GS24653: Heating jacket and controller for A10 as Cell P/N GS24654: Heating jacket and controller for A20 gas Cell The Atmos heating jacket is designed to fit and close around the metal body of the Atmos gas cell. The heating jacket overall is constructed using two inner C-section metal clamping bodies that are hinged together with one of the C-sections fitted with heating cartridges.
Page 5
4000 series temperature controller via the power cable lead using a six-way screw plug fixing. Power to the Atmos heating jacket is supplied by a dedicated 4000 series temperature controller. A separate user instruction manual for the 4000 Series temperature controller is supplied for use with the Atmos heating jacket instructions.
Before use of the Atmos heating jacket and controller with your own specific Atmos gas cell, please follow the safety advice for use of the Atmos gas cell itself as found from the Safety Considerations section in its own user instruction manual (2I-24602-3).
Page 7
Atmos Heating Jacket In use of an Atmos gas cell with its own heating jacket for the study of gaseous/vapour sample types, there is an increased risk associated with operation at elevated temperatures up to 200°C. Parts of the gas cell system will...
Page 8
200-210°C. Warning: The appropriate Atmos heating jacket must be fitted to its dedicated size of Atmos gas cell before it is heated. Warning: Do not move or touch the heating jacket on the Atmos gas cell whilst it is operating above ambient temperature conditions.
Page 9
(P/N GS24161), or to monitor the pressure to prevent for any over-pressurisation to the system. (Please check information found on pages 28 to 32 for Section 7), of the Atmos gas cell instruction manual 2I-24602-3. Important Notes For Safety In Operation...
Page 10
User Manual As a safety precaution the thermal fuse will blow to stop overheating (thermal run-away) of the Atmos heating jacket and gas cell when the temperature of the inner heating jacket (1) exceeds 210°C. Warning: In the event of a blown thermal fuse, the Atmos heating jacket with its 4000 Series temperature controller should be returned to Specac for repair.
GS24651 (for A2.5 gas cell), GS24652 (for A5 gas cell), GS24653 (for A10 gas cell) and GS24654 (for A20 gas cell), have the following specifications. A table for the Atmos heating jackets with their dimensions and as comparative images for their size are shown below. Atmos Heating...
Page 12
User Manual The images of the Atmos heating jackets as fitted to their corresponding size of Atmos gas cells have been shown for their overall height dimension from the optical axis (light beam) centerline reference.
Page 13
GS24652 – A5 Gas Cell 0.85 GS24653 – A10 Gas Cell GS24654 – A20 Gas Cell Specifications of Atmos Heating Jackets GS24651, GS24652, GS24653 and GS24654 Operating Temperature range: ambient to 200°C. Thermocouple: K Type (Chromel-Alumel). Heating rate (metal body): 90 mins ambient to 200°C (Or: 2°C / min).
Page 14
User Manual The configuration for a 4000 series temperature controller for the specific mains voltage supply is tabulated below. Voltage Frequency Max Power Fuse Rating Fuse Type (Volts) (Hertz) (Watts) (Amperes) The insulation rating of external circuits (appropriate for single fault condition) = basic insulation and protective (earth) bonding.
Atmos Heating Jacket 4. Checklist On receipt of your Atmos heating jacket and 4000 series temperature controller system please check that the following have been supplied: • Atmos heating jacket for appropriate A2.5, A5, A10 or A20 Atmos gas cell.
The inner heating jacket assembly (1) consists of two “C” section metal bodied parts that are hinged together allowing for the heating jacket to be opened and to fit around the metal body (28) of the Atmos gas cell. When fitted around the metal body (28), the hinged C-section parts of the inner heating jacket (1) are closed together and held securely shut by a tab clip bracket mechanism (2).
Page 17
The outer insulating jacket (3) consists of a heat resistant fabric casing surrounding the metal inner heating jacket (1). When the inner heating jacket (1) is closed to fit around the Atmos gas cell metal body (28), the outer insulating jacket (3) can be placed to cover and envelop the heating jacket (1).
Page 18
Fitting the Inner Heating Jacket (1) to the Atmos Gas Cell Having removed the Atmos gas cell from any Lever-lock baseplate mount fixing, position the Atmos gas cell onto a level workspace area. With the outer insulating jacket (3) removed, open the inner heating jacket assembly (1) (as seen in Fig 1.) by undoing the clip tab bracket...
Page 19
With the inner heating jacket (1) parts opened as wide as possible, carefully lift the heating jacket assembly (1) to lower over and surround the metal body (28) of the Atmos gas cell. The underside edge of the heating jacket assembly (3) should come to rest on the top surface of the lower screw fixing flange ring for the Atmos gas cell body (28).
Page 20
(24) of the optical unit for the Atmos gas cell. (See Fig 3.) The completed, correct fitting of the outer insulating jacket (3) to an Atmos A5 gas cell is shown as Fig 4.) Fig 4. Heating Jacket Fitted To A5 Atmos Gas Cell...
Page 21
Preparing for Operation With the Atmos heating jacket (1) and insulating jacket (3) correctly fitted and positioned over the Atmos gas cell, the whole gas cell and jacket assembly can now be fitted to a Lever-lock baseplate as installed in the spectrometer.
Page 22
“cleanest” reference conditions. N or any type of Infrared “inactive” gas would suffice, or even operating the gas cell under a vacuum. Ideally, when not in use, the Atmos gas cell should be stored under inert conditions to prevent environmental contamination.
Atmos Heating Jacket 6. Operating Parameters for Heating Jackets The Atmos Gas Cell Heating Jacket is provided with its own dedicated 4000 series temperature controller under the specific part numbers; GS24651 (A2.5 size gas cell), GS24652 (A5 size gas cell), GS24653 (A10 size gas cell) and GS24654 (A20 size gas cell).
Page 24
User Manual Displayable Parameters for Atmos A2.5 Gas Cell Heating Jacket P/N GS24651 with WEST 6100+ (4000 Series) Controller Parameter Parameter Name Parameter Factory Display (In Set Value Green) FiLt Input Filter Time Constant OFFS Process Variable Offset Primary (Heat) Output...
Page 25
Atmos Heating Jacket Displayable Parameters for Atmos A5 Gas Cell Heating Jacket P/N GS24652 with WEST 6100+ (4000 Series) Controller Parameter Parameter Name Parameter Factory Display (In Set Value Green) FiLt Input Filter Time Constant OFFS Process Variable Offset Primary (Heat) Output...
Page 26
User Manual Displayable Parameters for Atmos A10 Gas Cell Heating Jacket P/N GS24653 with WEST 6100+ (4000 Series) Controllers Parameter Parameter Name Parameter Factory Display (In Set Value Green) FiLt Input Filter Time Constant OFFS Process Variable Offset Primary (Heat) Output...
Page 27
Atmos Heating Jacket Displayable Parameters for Atmos A20 Gas Cell Heating Jacket P/N GS24654 with WEST 6100+ (4000 Series) Controllers Parameter Parameter Name Parameter Factory Display (In Set Value Green) FiLt Input Filter Time Constant OFFS Process Variable Offset Primary (Heat) Output...
User Manual 7. Cleaning/Care of the Atmos Heating Jacket After use of the Atmos gas cell fitted with a heating jacket it is recommended that the following procedures are adopted to properly care for the equipment. Flush the Atmos gas cell with dry N (Nitrogen) gas to remove any residual contaminants.
Page 29
Use a water dampened cloth when powered down only. If any parts of the Atmos gas cell such as the body, windows or mirrors require cleaning, then suitable solvents may be water, methanol, ethanol and acetone. When cleaning any of the mirrors (in the gas cell and in...
GS24627 ZnSe windows (pair) for Atmos gas cell. GS24628 CaF2 windows (pair) for Atmos gas cell. GS24625 Essential Spares Kit for Atmos A2.5 and A5 gas cells. GS24626 Essential Spares Kit for Atmos A10 and A20 gas cells. GS24160 Low or high pressure gauge kit.
Please use the table below to fill in the serial number information of the Atmos heating jacket and 4000 Series controller equipment received. If you need to contact Specac for any issues regarding your Atmos heating jacket, it may be necessary to provide the serial number of the item to identify for replacement parts.
(6) Tab handle of tab clip bracket (2). (7) Clip part of tab clip bracket (2). (8) Catch part of tab clip bracket (2). (24) Top section of optical unit of Atmos gas cell. (28) Atmos gas cell metal body assembly.
Page 36
User Manual Notes For Use Of Atmos Gas Cell...
Page 37
Fax: 061 726 6550 email: info@portmann-instruments.ch Germany Website:www.portmann-instruments.ch L.O.T. - Oriel GmbH & Co, KG - Darmstadt Tel: 06151 88060 SPECAC INC. Fax: 06151 880689 414 Commerce Drive email:info@LOT-Oriel.de Suite 175, Website: www.LOT-Oriel.com/de Fort Washington, PA 19034, USA...