Golden Gate™ High Temperature Heated Diamond Top Plate 1. Introduction The Golden Gate™ with High Temperature Heated Diamond ATR Top Plate (GS10642) is one of a series of top plates which can be used on the optical base assembly of the Golden Gate™ ATR system. The High Temperature Heated Diamond ATR Top Plate is operable across a temperature range from ambient to 300°C using the temperature controller.
User Manual 2. Safety Considerations List of Safety Symbols Safety Symbol Meaning General Caution. (Reference ISO 7000-0434B, 2004-01) Caution – Hot Surface. Reference IEC 60417-5041, 2002-10) Caution – Possibility of Electric Shock Indoor Use Only Electrical Safety Warning: The temperature controllers supplied with the Golden Gate accessory are of Class 1 (earthed) construction and must be connected to an earthed mains socket outlet.
Page 7
Golden Gate™ High Temperature Heated Diamond Top Plate General Operational Safety Use Caution: Always follow local laboratory safety protocols and procedures when using potential toxic or flammable substances with this equipment. The equipment is intended for use by suitably trained personnel only. Caution: For sample analysis of potential flammable liquids the puck is designed to operate with less than 0.1 ml in volume.
User Manual 3. Specification The Golden Gate™ High Temperature Heated Diamond ATR Top Plate is provided with its own dedicated Temperature Controller. A separate manual is supplied for specific operation of the Temperature Controller. This device is designed to use in a controlled laboratory environment only.
Page 9
Golden Gate™ High Temperature Heated Diamond Top Plate Golden Gate™ High Temperature Heated Diamond ATR Top Plate Specification with Temperature Controller Temperature range: Ambient to 300°C maximum. Thermocouple type: K Type (Chromel-Alumel). Heating rate (puck): 6 mins, 20°C to 300°C (≈45°C / min). Cooling rate: 25°C/min from 300°C.
User Manual 4. Unpacking and Checklist On receipt of the equipment please check that the following have been supplied: Golden Gate™ High Temperature Heated Diamond ATR Top Plate Golden Gate™ Optical unit with choice of ZnSe or KRS-5 lenses and appropriate Benchmark baseplate (if GS10642 ordered) ...
Opening of the compartment lid will allow natural heat convection to prevent the unit from overheating. However, if the sample compartment lid must be kept closed, Specac recommend that a purging gas of N is used through the sample compartment.
Page 12
Fax: 061 726 6550 email: info@portmann-instruments.ch Germany Website:www.portmann-instruments.ch L.O.T. - Oriel GmbH & Co, KG - Darmstadt Tel: 06151 88060 SPECAC INC. Fax: 06151 880689 414 Commerce Drive email:info@LOT-Oriel.de Suite 175, Website: www.LOT-Oriel.com/de Fort Washington, PA 19034, USA...
Page 13
Brilliant Spectroscopy™ www.specac.com SPECAC INC. SPECAC LTD. 414 Commerce Drive Science and Innovation Centre, Suite 175, Halo Business Park, Fort Washington, Cray Avenue, PA 19034, USA Orpington, Tel: 215 793 4044 Kent BR5 3FQ Fax: 215 793 4011 Tel:+44 (0) 1689 873134...