Error Correction

2. Camera light source:

Depending on manufacturer and type, the light controller reacts slowly or not at all.
Reduce the light intensity.
The connection cable from the light source controller to the camera amplification is
missing or defective.
The light regulator is defective.

3. Monitor:

If the brightness or contrast are set incorrectly, the image will appear to too bright or flat,
or black regions will appear gray.
Note when switching on the OSD (Operating Screen Display) for TFT Monitors:
18" Monitor
To block or unblock the OSD menu, quickly press (within 1 sec.) the following key combi-
nations: