Section 1
Installation Environment
Bench Space Needed
Total Height, tubes and block
Ambient Operating
Temperature Range
Altitude
Relative Humidity
Shelf Weight
Ventilation Needed
Safety
EN61010 Pollution Degree
EN61010 Installation Category
Micro Dist Heating Block
AC Input Frequency
Maximum Active Current (A)
Rating
Fuse Rating
Block Operating
Temperature Range
Controller Setpoint Error
Controller Precision
Hole-to-Hole Temperature
Variation
Sample Capacity
Micro Dist Tubes
Sample Size
Tube Material
Compliance
Certifications
Specifications
Specifications are subject to change without notice.
1.5 m (4 ft)
33 cm (13 inches)
5–40 °C
Up to 2 Km (6800 ft) above sea level
Up to 92.5%, non-condensing
20 kg (45 lb)
Ordinary air-conditioned lab. Provide at least 6 inches distance between the Micro Dist
block and all other vertical surfaces.
Laboratory 61010-1, 61010-2-010 (UL, CSA, EN Safety Standards)
II
II
100–115 V, 50/60 Hz
230 V, 50/60 Hz
100 V, 9 A
115 V, 11 A
230 V, 5 A
100–115 VAC Model: 15 A slow-blow, 125 VAC
230 VAC Model: 6.3 A slow-blow, 250 VAC
120 °C typical; 30–135 °C, L-type thermocouple, PID variable cycle controller
<1% of setpoint temperature
0.5 °C
<0.5 °C
21 Micro Dist™ Tubes
6.0 mL capacity
Analytical-grade pure polypropylene
100–115 V model: approved to UL and CSA safety standards (cETLus mark);
230 V model: CE certified by Hach
5