GREISINGER G 1107 Quick Reference Manual

GREISINGER G 1107 Quick Reference Manual

Fine manometer / manometer
Hide thumbs Also See for G 1107:

Advertisement

Quick Links

Quick reference guide
G 1107 / G 1113
EN
Fine manometer | manometer

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G 1107 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GREISINGER G 1107

  • Page 1 Quick reference guide G 1107 / G 1113 Fine manometer | manometer...
  • Page 2: Table Of Contents

    Measurement Basics.............. 34 Special functions................ 34 5.1.1 NVLL Tare function ................ 35 5.1.2 F,NE High-resolution measurement with 0.1 Pa (G 1107).... 35 5.1.3 F,NE High-resolution measurement with 1 Pa (G 1113).... 36 5.1.4 AVR 0:02 / AVR 0:05 / AVR 0: 10............ 36 Operation and maintenance ............ 38 Operating and maintenance notices .......... 38...
  • Page 3 Table of contents 6.2.2 Changing battery................ 38 Error and system messages ............ 39 Technical data ................ 40 Service...................  44 Manufacturer................... 44 B-H88.0.2X.DKZ-1.0 25 / 44...
  • Page 4: About This Documentation

    1 | About this documentation G 1107 / G 1113 1 About this documentation 1.1 Purpose of the document – This document is intended as a quick reference option. – It does not replace the operating manual. – For this reason, read the operating manual before operating the product for the first time.
  • Page 5: Safety

    G 1107 / G 1113 Safety | 2 2 Safety 2.1 Explanation of safety symbols DANGER This symbol warns of imminent danger, which can result in death, severe bodily in- jury, or severe property damage in case of non-observance. CAUTION This symbol warns of potential dangers or harmful situations, which can cause dam- age to the device or to the environment in case of non-observance.
  • Page 6: Safety Instructions

    2 | Safety G 1107 / G 1113 2.3 Safety instructions NOTE This product does not belong in children's hands! 2.4 Intended use The device is designed as a manometer and measures even the smallest pressure differential pressures of up to ±200 hPa with a maximum resolution of up to 0.1 Pa in air or in non-corrosive/non-ionising gases between the two pressure connections.
  • Page 7: The Product At A Glance

    G 1107 / G 1113 The product at a glance | 3 3 The product at a glance 3.1 The G 1100 manometer series Top view Front view LCD Display Connection variants 3.2 Display elements Display Battery indicator Evaluation of the battery status...
  • Page 8: Connections

    3 | The product at a glance G 1107 / G 1113 Up / Down button Press briefly Display of the min/max value Change value of the selected parameter Long press Reset the min/max value of the current measure- ment...
  • Page 9: Operation

    G 1107 / G 1113 Operation | 4 4 Operation 4.1 Opening the configuration menu 1. Press the Function key for 2 seconds to open the Configuration menu. 2. (ONF appears in the display. Release the Function key. Parameter Values...
  • Page 10 4 | Operation G 1107 / G 1113 Measuring rate Selection of the measurement speed RATE Slow Fast FAST Resolution Selection display resolution RANG Automatic switchover AVTO G 1107 Adjusted to the highest value, e.g. -200.0 .. +200.0 Adjusted to the lowest value, e.g. -20.00 .. +20.00 G 1113 Adjusted to the highest value, e.g.
  • Page 11: Open The Adjustment Menu

    G 1107 / G 1113 Operation | 4 Backlight L,TE Backlight deactivated 0:15 0:30 1:00 4:00 Automatic shut-off of the backlight after a selected time in minutes and seconds, during which no but- tons have been pressed No automatic shut off of the backlight...
  • Page 12: Measurement Basics

    5 | Measurement Basics G 1107 / G 1113 5 Measurement Basics 5.1 Special functions With the special functions that can be selected via the Configuration menu, the device can be optimised for special measuring tasks. After it is switched on, the device starts up in standard measuring mode, the relevant special function is started by pressing and holding the Function key for 4 s.
  • Page 13: Nvll Tare Function

    The tare function is independent of the zero point correction accessible via the set- tings menu. 5.1.2 F,NE High-resolution measurement with 0.1 Pa (G 1107) High-resolution measurement for the finest adjustment work, 4 Pascal Test (test of chimney draft with living-space-independent single combustion) and many other finely-adjusted pressure applications.
  • Page 14: F,Ne High-Resolution Measurement With 1 Pa (G 1113)

    5 | Measurement Basics G 1107 / G 1113 5.1.3 F,NE High-resolution measurement with 1 Pa (G 1113) High-resolution measurement for the finest adjustment work and many other applic- ations with the finest adjustment of pressure. In the Configuration menu, the special function FVNC F,NE has been selected.
  • Page 15 G 1107 / G 1113 Measurement Basics | 5 Heavily fluctuating values arise particularly with dynamic pressure/compression measurements in chimney draft tests of forced-air burners and, consequently, con- ventional electronic manometers are not adequate for task. This special function op- timises the device for this application purpose.
  • Page 16: Operation And Maintenance

    6 | Measurement Basics G 1107 / G 1113 6 Operation and maintenance 6.1 Operating and maintenance notices NOTE Pressure connections must be protected from soiling. 6.2 Battery 6.2.1 Battery indicator For additional information, refer to the operating manual! 6.2.2 Changing battery...
  • Page 17: Error And System Messages

    G 1107 / G 1113 Error and system messages | 7 7 Error and system messages Display Meaning Possible causes Remedy Calculation not pos- Measurement data Waiting for data collection ---- sible acquisition is run- ning No display, Battery depleted...
  • Page 18: Technical Data

    8 | Technical data G 1107 / G 1113 8 Technical data G 1107 Measuring range Measuring range (Hi) Measuring range (Lo) -2000 .. +2000 Pa -200.0 .. +200.0 hPa -20.00 .. +20.00 hPa (mbar) (mbar) -2.900 .. +2.900 PSI -150.0 ..
  • Page 19 G 1107 / G 1113 Technical data | 8 Operating conditions -20 to +50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.) Storage temperature -20 to +70 °C Current 2*AA battery (included in the scope of delivery) supply Current consump- approx.
  • Page 20 8 | Technical data G 1107 / G 1113 G 1113 Measuring range Measuring range (Hi) Measuring range (Lo) -2000 .. +2000 hPa -200.0 .. +200.0 hPa (mbar) (mbar) -2.000 .. +2.000 bar -29.00 .. +29.00 PSI -1500 .. +1500 mmHg -200.0 ..
  • Page 21 G 1107 / G 1113 Technical data | 8 Storage temperature -20 to +70 °C Current 2*AA battery (included in the scope of delivery) supply Current require- approx. 1 mA (slow measurement SLO) ment/ Operating time approx. 3000 h Battery life...
  • Page 22: Service

    9 | Service G 1107 / G 1113 9 Service 9.1 Manufacturer If you have any questions, please do not hesitate to contact us: Contact GHM Messtechnik GmbH GHM GROUP - Greisinger Hans-Sachs-Str. 26 93128 Regenstauf | GERMANY Email: info@greisinger.de | www.greisinger.de WEEE reg.

This manual is also suitable for:

G 1113

Table of Contents