Scope of Delivery

2.4

Optional Accessories

2.5
8
The LabTouch-aw instrument is delivered as followed:
LabTouch-aw:
Instrument LabTouch
Power supply with different insertion plugs
4 pcs of humidity standards : SAL-T: 33, 58, 75 and 84%rH
40 pcs standardized disposable sample dishes (ePW)
Operating manual in English and German on a USB memory stick
Factory calibration certificate
SD-card 8 GB
Tension ring
5 pcs pre-filter (white)
USB and RS-232 cables
Mechanical filter
The white pre-filter provides mechanical protection of the measurement cell
e.g. in case the sample cup is overfilled, sample will stick on the filter instead
of the measurement cell.
This filter should be replaced when it is dirty or at least on a yearly base,
otherwise it could influence the measurement. 5 units are included in the
delivery of the instrument.
Chemical protection filter
Some samples do have volatiles components / ingredients thus the measuring
cell has to be protected from damaging, gaseous substances. This is done by
various filter systems which are provided by Novasina. To select an appropri-
ate filter system, consult the Novasina protection filter systems data sheet or
contact your local Novasina distribution partner.
Thanks to these filter systems, the lifetime of the measurement cell can be
extended considerably. As most of the filter material do absorb the volatile
component and get saturated by time, a regular check is mandatory to avoid
a contamination of the sensor by an ineffective, saturated filter system