Sensor Calibration

How to Handle SAL-T Standards Correctly
8.1
stable protocol" is set to "Off".
Transmittance to the Novalog MC software can be done by selecting "PC" or
"SD+PC". This way, a automatic print-out of the protocol on any connected
PC printer (incl. pdf) can be achieved . "PC" or "SD+PC" are only displayed
if the the instrument has been licensed for NovalogMC.
Manual saving of a protocol
A protocol of the current values ("Actual protocol") can be saved anytime
and even during an on-going measurement on the SD-card by pressing the
printer symbol on the main menu screen. Just select "Actual to SD-Card"
on the appearing pop-up window.
A "Stable protocol" can only be stored this way, if stable values are present,
respectively if a measurement value had been considered as stable. If these
pre-requisites are fulfilled, a pop-up appears after pressing the print button
and you have to select "Stable to SD-card" to initiate the saving process.
The protocols can be transfered directly to a PC if the instrument has been
recognized by the optional NovalogMC software by selecting "Actual to PC"
respectively "Stable to PC" after pressing the printer symbol
"Actual to PC" and "Stable to PC" are not displayed without a connection
to NovalogMC
The electrolytic sensor can be verified at any time by using the Novasina SAL-T
standards which are included in the delivery or which have been bought through
the local Novasina distributor. These reusable standards generate defined and
reproducible humidity, respectively aw-values. If verification fails, a calibration
can be performed in an easy way.
Note:The calibration values will be stored in a small memory located on the
sensor board. It is not required that a calibration is performed at all available
calibration points. Perform a two-point calibration with one point below and one
above of the normal measurement range (bracket calibration).
For example: Your samples are always around 0.700aw. It is sufficient if you
use the 58%rH and the 75.%rH SAL-T standards for calibration. If verification
is requested, calibrate at 58%rH and 84% rH and use the 75%rH SAL-T stan-
dards for verification.
The Novasina SAL-T standards are not labeled with a due date. The shelf-life
itself depends on the handling and storage conditions. Experiences in the field
have proven that those standards can be good for 3-5 years if handled correctly.
To achieve the best possible shelf-life, please
Shake all SAL-T standards twice a month for one minute, nevertheless if
you have used them or not. There is no need to take it out of the plastic
container.
Always store the standards in the plastic container (close tightly) and not
exposed to direct sunlight.
Do not expose the salt tablet to environmental air for longer than necessa-
ry; thus, put it back if calibration has been finished at the particular point.