Front Panel Keys (cont.)
E
Loads a newly keyed-in output value showing on
the Control Display into the Calibrator. If you press
E without identifying the units for the entry,
the Calibrator keeps the units (except Hz) that
were last used. This allows you to alter values for
a function without keying in the units every time.
The multipliers are not retained, for example, you
can enter 1 mV, and then later simply enter 10 to
obtain 10V.
In addition to these functions, the E key
recalls the currently programmed reference value
during error mode operation.
D V A Q H
(Output Function Keys)
These keys select the output function as follows:
D
Decibels relative to 1 mW
V
Voltage
A
Current
Q
Resistance
H
Frequency
When a value, then H are selected, the
Calibrator automatically switches to ac. To change
back to dc, enter 0 H or enter a signed (+ or -
amplitude).
U m K M
(Multiplier Keys)
These keys select output value multipliers. For
example, if you enter 33, then m, then V, then
ENTER, the Calibrator output value is 33 mV.
U
micro (10
m
milli (10
K
kilo (10
M
Mega(10
-6
or 0.000001)
-3
or 0.001
3
or 1000)
6
or 1,000,000)