Chapter 12 Calibration Tmx428 Series; Belt Speed Calibration Procedure (Calibration Program) - Trackmaster TMX428 Service Manual

Medical treadmill
Table of Contents

Advertisement

TRACKMASTER
Calibration TMX428 Series

Belt Speed Calibration Procedure (Calibration Program)

Tools Required
Calibration Program (Full Vision Treadmill PS New Cal)
Stop watch or watch with a second hand
Chalk or some other temporary marking device
Computer or laptop
Serial Communication Cable
USB Communication Cable
1. Plug treadmill to power supply.
2. Connect computer or laptop to treadmill RS232 port or USB mini port.
3. Turn ON Treadmill.
4. Open Full Vision Treadmill PS New Cal software.
5. Select Comport and press Connect.
6. Verify communications is receiving and sending data.
7. Remove all tools from running belt and stand clear of running surface.
8. Select
9. Observe: Running belt will pause for approximately 8 seconds and slowly increase speed
establishing multiple calibration points.
10. When calibration routing is complete and successful (
automatically update Cal Speed Fields. When calibration routine fails
11. At this point running belt calibration is complete.
12. To set MPH versus KPH, position dip switch 1 when in the "OFF" position, units are set to
English, and in the "ON" position, units are set to Metric.
13. Select
14. To change speed select
15. Measure belt with MPH/KPH gauge or using "Verifying Belt Speed Calibration (in the
field)". Verify within specification +/- 2%.
16. Repeat process for 2, 3, 5, 7, & 12 MPH or 4, 6, 8, 12 & 19 KPH.
17. If not with tolerance, verify
recalibrate belt.
18. Repeat steps 13 thru 16 until within tolerance.
19.
317-160-406 Rev 4
May 2021
Ch. 12
), consult factory for assistance.
or
TMX428 TMX428CP & TMX58 Service Manual
Removal and Replacement of Components
and the running belt will start at 0.1 MPH.
and type desired speed.
and
12
) value will
(
, then
117

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tmx428cp

Table of Contents