Specifications - Olympus TH4 Instructions Manual

Halogen lamp power supply unit
Table of Contents

Advertisement

SPECIFICATIONS

Item
Input rating
TH4-100
TH4-200
Output rating
Ventilation
Dimensions
Weight
Operating environment
Specification
100-120 V
, 50/60 Hz, 1.8 A
200-240 V
, 50/60 Hz, 0.9 A
0.9 to 12.6 V DC, 8.4A
Natural
# Be sure to leave at least 10 cm (4 in.) of free spaces
cooling
between the walls or other solid objects.
system
75 mm (W) x 125 m (H) x 200 mm (D)
Approx. 2.3 kg
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F),
decreasing linearly through 70% at 34°C (93°F), 60% at 37°C
(99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: Not to exceed ±10% of the nominal
voltage.
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation (overvoltage) category: II (in accordance with IEC60664-1)
· Transient overvoltage: 2500 V
· Temporary overvoltage:
TH4-100: 1320 V (up to 5 sec); 370 V (longer than 5 sec.)
TH4-200: 1440 V (up to 5 sec); 490 V (longer than 5 sec.)
TH4
8