Contact corrosion due to moisture.
Damage to the product.
▶ Remove the product from the steam sterilization immediately after the
sterilization cycle.
The KaVo medical device has a maximum temperature resistance of up to
138 ℃ (280.4 °F).
Sterilization parameters:
Select a suitable process from the following sterilization processes (depending
on the available sterilizer):
▪ Sterilizer with triple pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
– Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- drying time 30 minutes or alternatively
- at least 60 minutes at 121°C (250 °F)
- Drying time: 15 min.
▶ Remove the medical device immediately after the completion of the steril-
ization cycle from the sterilizer.
▶ Use in accordance with the manufacturer's Instructions for Use.
7.7 Storage
Processed products must be stored, protected from bacteria, to the extent pos-
sible, and dust, in a dry, dark, cool room.
Note
Comply with the expiration date of the sterilized items.
7 Processing steps in accordance with ISO 17664 | 7.7 Storage
NOTICE
25 / 30
Instructions for use elements™ 8:1