Operating Modes Of The Y Amplifier - Hameg HM 604 Manual

Hide thumbs Also See for HM 604:
Table of Contents

Advertisement

defined area of influence on the waveform shape (see Fig.),
offering the added advantage of being able to 'straighten
out' waveform aberrations near the leading edge.
Adjustment points of the probes
alters the middle frequencies
alters the leading edge
T,:
alters the lower frequencies
After completion of the HF-adjustment, the signal
amplitude displayed on the CRT screen should have the
same value as during the 1 kHz adjustment.
Probes other than those mentioned above, normally have a
larger tip diameter and may not fit into the calibrator out-
puts Whilst it is not difficult for an experienced operator to
build a suitable adapter, it should be pointed out that most
of these probes have a slower risetime with the effect that
the total bandwidth of scope together with probe may fall
far below that of the HM604. Furthermore, the HF-adjust-
ment feature is nearly always missing so that waveform dis-
tortion can not be entirely excluded.
incorrect
Adjustment
1 MHz
Subject to change without notice
HZ52
incorrect
correct
The adjustment sequence must be followed in the order
described, i.e. first at
frequencies should not be used for timebase calibrations.
The pulse duty cycle deviates from 1 : 1 ratio.
Prerequisites for precise and easy probe adjustments, as
well as checks of deflection coefficients, are straight hori-
zontal pulse tops, calibrated pulse amplitude, and zero-
CAL.
potential at the pulse base. Frequency and duty cycle are
relatively uncritical.
response,
fast pulse risetimes and low-impedance
generator outputs are of particular importance.
Providing these essential features, as well as switch-select-
able output-frequencies, the calibrator of the HM 604 can,
under certain conditions, replace expensive squarewave
generators when testing or compensating wideband-
attenuators or -amplifiers. In such a case, the input of an
appropriate circuit will be connected to one of the CAL.-out-
puts via a suitable probe.
The voltage provided at a high-impedance input (I MSJ II 15-
1 =
(10:
Suitable probes are HZ51, 52, 53, and 54.
For low-impedance inputs (e.g. 50 Q), a 1: 1 probe can be
employed which, however, must be fully terminated with a
and HZ54. The latter must be switched to the 1: 1 position,
and the HF-trimmer in the connecting box turned fullycoun-
terclockwise.
When connected to the
HZ50, this arrangement will provide approx. 40mV,, at
The voltages given here will have larger tolerances than 1 %
since operation of a 1: 1 probe together with a 5OQ load is
very uncommon.
Using the 2V CAL. socket under similar conditions is only
possible with the HZ54 probe. The potential obtained at the
twice the risetime. Accurate readings of the available input
voltage can be shown directly on the HM604 when con-
necting a 5OQ through-termination between the BNC plug
of the probe and the input of the oscilloscope.

Operating Modes of the Y Amplifier

The required operating modes are selected on three
pushbuttons located in the Y-Section. For
all pushbuttons should be in the out position, the instru-
ment is then operating on
1
1
MHz. The calibrator
For interpretations of transient
1 =
also 20mV,, from 2V output).
socket, and using the
Mono
Channel/only.
operation
604

Advertisement

Table of Contents
loading

Table of Contents