Philips AVENT Premium Plus SCF394 Manual page 28

Single/double electric breast pump
Hide thumbs Also See for AVENT Premium Plus SCF394:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
English
28
External dimensions
Device classification
Materials
Service life
Mode of operation
Ingress protection
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying
tables. Only use this Philips Avent single/double electric breast pump Premium Plus in combination
with the adapter supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Premium Plus is intended for use in the
electromagnetic environment listed below. The user of the device should ensure that it is used in such
an environment to prevent improper operation.
Single electric breast pump
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Applicable classifications:
Degree of protection against electric shock:
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical
system according to IEC 60601-1.
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Double electric breast pump
Non-continuous operation
55 min on, 180 min off at 22 °C
(72 °F) (duty cycle 23 %)
20 min on, 90 min off at 40 °C
(104 °F) (duty cycle 18 %)
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in
establishments directly connected to a low
voltage power supply network which supplies
buildings used for domestic purposes.

Advertisement

Table of Contents
loading

Table of Contents