Output level accuracy (IQ
modulation relative to CW)
This test checks the output level accuracy of the IQ modulation inputs
relative to continuous waveform

Required equipment

Spectrum analyzer: Tektronix RSA5106B
Arbitrary function generator: Tektronix AFG3252C
Two BNC - BNC cables
Type-N (male) - Type-N (male) cable
1. Using a BNC to BNC cable, connect the AFG3252C (Channel 1)
to the Vector Modulation I input of the TSG (located on the rear
panel).
2. Using the second BNC to BNC cable, connect the AFG3252C
(Channel 2) to the Vector Modulation Q input of the TSG
(located on the rear panel).
3. Set the AFG3252C as follows:
a. Set phase on Channel 1 to 90, frequency to 1 MHz, and
amplitude to 1 Vp-p.
b. Set phase on Channel 2 to zero, frequency to 1 MHz,
and amplitude to 1 Vp-p.
c. Press Phase/Delay and then Align phase in the menu.
This aligns the phase between Channel 1 and Channel 2
to obtain the quadrature input for the TSG.
4. Set the TSG as follows:
a. Press and hold the Preset button to reset the
instrument.
b. Press the RF/LF button to access the output settings
menu.
c. Press the Freq button on the front panel.
d. Set the frequency to 2000 MHz (fc) using the general
knob or the number keys.
e. Select RF Ampt from the menu.
f.
Set the RF amplitude to 0 dBm using the general knob
or number keys.
g. Connect the output from the TSG to the input of the
spectrum analyzer.
h. Press the gray color
panel to access the modulation parameters menu.
i.
Set the Source to Ext.
j.
Select RF Output from the menu to turn the RF output
to ON (text will turn blue).
5. Set the spectrum analyzer as follows:
a. Reset the instrument.
b. Set the center frequency to 2000 MHz and the span to
5 MHz.
c. Set markers to peak search.
Mod button on the front

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TSG4100A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Tektronix TSG4100A

  • Page 1 CW) relative to continuous waveform Required equipment Spectrum analyzer: Tektronix RSA5106B Arbitrary function generator: Tektronix AFG3252C Two BNC - BNC cables Type-N (male) - Type-N (male) cable 1. Using a BNC to BNC cable, connect the AFG3252C (Channel 1) to the Vector Modulation I input of the TSG (located on the rear panel).
  • Page 2 d. Write down the peak value and note it is = A for later reference. 6. Press the Mod button on the TSG front panel to turn modulation on. Notice that MODON indicator will appear in yellow in the top right portion of the screen. 7.

Table of Contents