Step 6
Insert the HEPA filter (metal side downwards) followed by the
carbon filter into the filter cassette. To place the filter cassette
into the top assembly move it upwards into the opening and
press gentle until the handles locks into place.
Finally connect the system to the power supply by plugging the power plug  from the back panel 
into the local power outlet.
The system is now ready for use.
AC power outlets
The 4 AC power outlets serve as power supplies for devices used inside the PCR chamber.
The sum of the maximum power ratings of devices connected to these power outlets
must not exceed the values specified for the PCR Workstation.
OPERATING THE PCR WORKSTATION
Use of PCR equipment inside the workstation
Shelves are provided for placement of small tools for storage and decontamination.
Power outlets allow use of equipment for PCR experiments inside the PCR chamber. The equipment
can be decontaminated between experiments.
Timing of the UV decontamination can be programmed for time-controlled every day decontamination.
NOTE: When operating equipment plugged into the outlets, combined Amps total is not to
exceed 6 Amps.
Operating the UV and white lights
The UV light is typically operated when no samples are inside the chamber. The chamber is
decontaminated and DNA as possible contamination is inactivated.
Timer
The default setting is 30 minutes adjustable at 5 minutes increments down to a minimum of 5 minutes.
When the front panel is opened, the ultraviolet light will automatically shut off!
Note:
Do not attempt to perform PCR procedures with the germicidal lamp on. The germicidal lamp is
used to decontaminate the chamber between experiments.