6.3 Check Mode

Check mode consists of three modes: Data adjustment mode, function setting mode and
basic setting mode.
1. Entering the check mode
GX- 203A/303A/403A/603A/1003A/1603A
2002A/3002A/4002A/6002A/10002A
6001A/10001A
GF- 123A/203A/303A/403A/603A/1003A/1603A
1202A/2002A/3002A/4002A/6002A/10002A
6001A/10001A
Step 1.
Verify that the display is OFF.
Step 2.
Press and hold the RE-ZERO and PRINT keys and press the ON/OFF key.
Release the PRINT and ON/OFF keys while still holding the RE-ZERO key.
Immediately press the PRINT key twice. Perform this procedure within 2 seconds.
Step 3.
The software version will be displayed for about 1 second
Step 4.
The balance model type will be displayed GX203A . (The model type displayed
depends on each actual model.)
Step 5.
All of the display segments will turn on.
GX-A/GF-A Maintenance manual
Page 28
p- X.XX
.