Certified medical device
BM-03 Baby breathing monitor

Advertisement

Table of Contents
loading

Summary of Contents for nanny BM-03

  • Page 1 Certified medical device BM-03 Baby breathing monitor...
  • Page 2: Table Of Contents

    Immunity requirements - input and output through the device’s cover from RF devices Prior to using the BM-03 Breathing Monitor, please read the instruc- tions for use and the conditions of its use carefully, as well as the general procedures of first aid and emergency care for children! If you have any questions about using the monitor, please refer to the contacts provided in this manual.
  • Page 3: Product Identification

    PRODUCT IDENTIFICATION The BM-03 Baby Breathing Monitor is a certifi ed Class IIb medical device that monitors a baby’s breathing. It cannot be used to restore vital functions and is not a therapeutic device. It does not replace proper childcare.
  • Page 4: Functions And Controls

    1. mechanical on/off button for the monitor 2. optical breathing/alarm signal 3. the NANNY logo works as a capacitive button for light control 4. speaker 5. the indicator light with the power-on symbol - signaling of laying down/...
  • Page 5: Supplementary Functions

    DETECTING WHETHER THE BABY HAS BEEN PLACED ON THE SENSOR PAD OR TAKEN OFF IT (SWITCH GUARD) When in standby mode the BM-03 Baby Breathing Monitor continuously evaluates signals from the pad and is thus the only device on the market today to point out that the baby was most likely placed on the sensor pad.
  • Page 6: Night Light

    NANNY logo (Fig. 5a), for 10 secs. After activation, the lamp flashes three times. The light function can be deactivated in the same way.
  • Page 7: Monitoring Room Temperature

    NANNY logo. Place your fingertip on the center of the NANNY logo and leave it attached for at least 1 second (Fig. 5b). There is no need to press the switch, just place your finger on the surface.
  • Page 8: Installation

    EN | Baby breathing monitor INSTALLATION Neither the sensor pad nor the control unit need to be specially cleaned or disinfected after unpacking. Before using the monitor, make sure that all parts are undamaged. The device does not require temperature adaptation in connection with its installation and subsequent repeated use.
  • Page 9 5. For your convenience, you can hold the control unit with various accessories: • a cot holder and Velcro - you can tighten the Velcro to the circumference of the upper edge of the cot; • holder and double-sided adhesive tape to place it on the side of furniture or another solid surface;...
  • Page 10: In Situ Monitor Usability Test

    EN | Baby breathing monitor IN SITU MONITOR USABILITY TEST We recommend performing the usability test daily, but at least when changing the location of the bed or monitor. • Make sure the green light flashes while the baby is in the cot. The green light reacts by flashing with the baby’s breathing or movements.
  • Page 11: Sensor Pad

    SENSOR PAD The sensor pad is included in the package. It can also be bought separately as a spare part labelled BM-03D. One sensor pad is enough until the baby is about 6 months old. If the baby starts to climb or move in the cot, the monitored area can be enlarged by attaching a second sensor pad.
  • Page 12: Life Span Of The Sensor Pad

    The life span of the sensor pad is set at 2 years, after which the pad must be replaced. As auxiliary data, the Nanny BM-03 breathing monitor also monitors the number of hours the monitor has been in operation - if it is likely the pad has exceeded the specifi...
  • Page 13: Alarm States

    ALARM STATES PHYSIOLOGICAL ALARMS Alarm state Priority Light indication Acoustic indication Low breathing rate Red LEDs flash in a circle around Alarm - continuous the mechanical button Number of inhalations/ex- High melody 80 dB/m +/- 5% halations less than 8/min. (2.5 Hz, 200 light/200 ms pause) (evaluated in active mode) Respiratory arrest...
  • Page 14: Operator Warning

    EN | Baby breathing monitor OPERATOR WARNING Event Light indication Acoustic indication The indicator light with the battery symbol Low battery flashes slowly and red No acoustic signal (100 ms light/2 s pause). The indicator light with the battery symbol Critically low battery flashes quickly and red No acoustic signal...
  • Page 15: Important Warning

    Use of the Nanny Breathing Monitor is not determined by the child’s age, but by its weight. The manufacturer recommends a Nanny breathing monitor for children from a min. weight of 1 Age of the baby kg and a max. weight of up to 15 kg. A heavier child may cause mechanical damage to the sensor.
  • Page 16 The manufacturer is responsible for the functionality of the BM-03 product if it is installed and used according to this ma- nual and its recommendations, otherwise it does not bear any Guarantee responsibility.
  • Page 17 PROBLEMS AND HOW TO SOLVE THEM child’s movement did not 1) Whilst breathing the reliably reach the sensor pad. • For very small babies, the sensor pad can be placed directly under the sheet (but always use an insulating pad to prevent liquids leaking into the pad).
  • Page 18: Technical Specifications

    EN | Baby breathing monitor TECHNICAL SPECIFICATIONS Power supply 3 V; 2 × 1.5 V alkaline battery type AA (LR6) Idle consumption 106 µA Consumption during an alarm 270 mA Low battery indicator voltage 2.46 V ± 0.15 V Critically low battery indicator voltage 2.2 V ±...
  • Page 19: Emc Of A Medical Device

    The product has been clinically evaluated and is registered by the Ministry of Health of the Czech Republic as a medical device class IIb. JABLOTRON ALARMS a.s. declares that the product BM-03 is designed and manufactured in accordance with the harmonization legislation of the European Union: Directive No.
  • Page 20: Immunity Requirements - Input And Output Through The Device's Cover

    EN | Baby breathing monitor 14.2 IMMUNITY REQUIREMENTS - INPUT AND OUTPUT THROUGH THE DEVICE’S COVER Immunity test levels Basic standard for EMC or Phenomenon test method Professional medical Home health care facilities ±8 kV or contact charge Electrostatic discharge IEC 61000-4-2 ±2 kV, ±4 kV, ±8 kV, ±...
  • Page 21 14.3 IMMUNITY REQUIREMENTS - INPUT AND OUTPUT THROUGH THE DEVICE’S COVER FROM RF DEVICES Test Maximum Immunity Band Distance frequency Service Modulation output test level Pulse modulation 380 to 390 TETRA 400 18 Hz FMc) ±deviation 5 kHz 130 to 470 GMRS 460 FRS 460 1 kHz sinusoidal waveform...