Spectral Irradiance (bare fiberoptic pad)*
*Using an Ohmeda Medical BiliBlanket
Meter II
Wavelength
LED module estimated life
Sound level
X-ray
Light box (W x H x L)
Light box weight (excluding fiberoptic pad)
Fiberoptic Pad weight
Fiberoptic light pad, small
Fiberoptic light pad, large
Fiberoptic cable length
IEC Type B equipment.
IEC Class 1 (continuous operation)
FDA Class II
The LED module life may vary when used in the actual clinical environment. Factors such as duty cycle and
*
ambient temperature may impact the life of the LED module. Measure the irradiance of the BiliSoft System and
replace the LED module when the system is below specifications.
9-2
Performance Specifications
Large Fiberoptic Pad – 49
25%) 9-point check
Small Fiberoptic Pad – 70
25%) 6-point check
Note: When the BiliSoft fiberoptic light pad is
inserted into a BiliSoft Pad Cover or BiliSoft
Nest, the nominal spectral irradiance is 35
µW•cm
•nm
2
430-490 nm (peak 440-460 nm)
Under continuous use, tested at room
*
temperature, a typical LED module will run for
approximately 8,000 to 10,000 hours before
the light intensity drops 25%.
< 44 dB(A) at 1 meter
X-ray compatible
Physical Specifications
16.5 x 21 x 16.5 cm
< 2.5 kg
< 1.1 kg
15 x 30 cm (light-emitting area)
25 x 30 cm (light-emitting area)
137 ± 5 cm
Regulatory Standards
Product certified to the following
standards:
EN60601-1; EN60601-1-2; EN60601-2-50; ISO
10993-5; ISO 10993-10; UL 60601-1; CSA
C22.2 No 601.1-M90; IEC 60601-1-8; BS EN
980; 16CFR Part 1632.6 (for BiliSoft Pad
Covers and BiliSoft Nests)
M1110147 Rev.
5
µW•cm
•nm
(large pad) and 50 µW•cm
-2
-1
(small pad)
-1
© 2007 Datex-Ohmeda. All rights reserved.
µW•cm
•nm
(+/-
-2
-1
•nm
(+/-
-2
-1
-

Hide quick links: