GE BiliSoft Operation And Maintenance Manual page 14

Phototherapy system
Hide thumbs Also See for BiliSoft:
Table of Contents

Advertisement

WARNING:
Water Loss: The radiant energy from phototherapy lights can increase a
patient's insensible water loss. Take appropriate measures to maintain
the patient's fluid balance while administering phototherapy.
WARNING:
Skin Temperature: Phototherapy light can affect the temperature in
thermoregulation devices (incubator, radiant warmers or heated
mattresses) and may raise the patient's body temperature. It is
recommended to use the incubator or warmer in skin controlled (servo)
mode. Always monitor the patient's temperature to avoid temperature
fluctuations during phototherapy.
WARNING:
Reflective Foils: Using reflective foils to increase the efficacy of
phototherapy may cause hazardous patient body temperatures.
WARNING:
Combustible Gases: Do not use the BiliSoft LED Phototherapy System in
the presence of gases that support combustion like oxygen, nitrous
oxide or other flammable or combustible anesthetics.
WARNING:
Flammable Solutions: Never use flammable solutions to clean the BiliSoft LED
Phototherapy System or any of its parts.
CAUTION:
Medical Electrical Equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
provided in the service manual.
Portable and mobile RF communications equipment can affect Medical
Electrical Equipment.
The BiliSoft LED Phototherapy System should not be used adjacent to or be
stacked on other equipment. If adjacent or stacked use is necessary, the
BiliSoft LED Phototherapy System should be observed to verify normal
operation in the configuration in which it will be used.
2-4
M1110147 Rev.
5
© 2007 Datex-Ohmeda. All rights reserved.