Reflected-Light Illumination Equipment; Change Of 6 V 30 W Halogen Lamp (Axiotech 30) - Zeiss Axiotech Operating Manual

Table of Contents

Advertisement

Carl Zeiss
2.12

Reflected-light illumination equipment

Fig. 2-28
Change of 6 V 30 W halogen lamp
Use a protective sleeve (2-31/3) to hold the bulb of the 6 V 30 W halogen lamp; if required,
clean the lamp bulb before its initial use with pure alcohol to prevent dirt from burning in.
2-22
1
4
2.12.1
Change of 6 V 30 W halogen lamp
(Axiotech 30)
• Switch off the On/Off switch (2-28/8) on the
microscope
stand
instrument cable (2-28/7) from the line.
• Allow hot halogen illuminator to cool down for
approx. 15 minutes after use.
• Unscrew hexagonal screws (2-28/6) using
SW 3 ball-headed screwdriver and remove lamp
cover (2-28/5).
• Press holding components (2-28/4) of lamp
insert (2-28/3) together and pull out lamp
insert.
• Remove new 6 V 30 W halogen lamp (2-28/1)
from the packaging and push it on the guiding
rails. Make sure that the lamp carrier plate
exactly fits the mounting and fixation pins.
• Attach lamp insert again by pressing the
holding components slightly together.
• Attach and screw on lamp cover using the
SW 3 ball-headed screwdriver.
• Connect instrument to the line and switch on
the On/Off switch on the stand.
Axiotech and Axiotech
and
disconnect
the
B 40-020 e 06/99