9 Accessories

Printer: Art. no 800-10 Model EPSON TM-T88V supplied with screened cable.
Clamp: Art. no 800-25.
Mobile stand: Art. no 400-45.
Basket: Art. no 400-85.
Hpm™ Reusable and autoclavable single and double cuffs, Hpm™ Disposable single and double
cuffs. Hpm™ disposable cuffs are sterilized and are only to be used according to hospital
procedures and regulations. The following codes are referred to as Hpm™ Disposable cuffs;
Hpm™ single disposable cuffs: 15-100, 15-110, 15-120, 15-130, 15-140, 15-150, 15-160, 15-
170 & 15-180.
Hpm™ single disposable cuffs conical: 75-140, 75-150, 75-160 & 75-170.
Hpm™ double disposable cuffs: 25-100, 25-110, 25-120, 25-140 & 25-160.
Hpm™ Tourniquet 800-series are only to be used together with cuffs marked according to the
CE standard.
We recommend that you only use Hpm™ CE-marked cuffs with Hpm™ Tourniquet 800-20 and
800-40.

10 Appendix EMC guidance

The following guidance information are required by EN 60601-1-2 (2007):
Warning
The tourniquet should not be used adjacent to or stacked with other equipment.
Warning
The use of any other printer than listed in chapter 9 may result in increased emisssions or
decreased immunity of the tourniquet.
Essential performance (as defined in EN60601-1-2)
 An alarm system should signal whenever the tourniquet inflated pressure is 30 mmHg
higher or 15 mmHg lower than the set pressure.
 In single fault condition the cuffs should be isolated using a block valve with a maximum
leakage of 1 mmHg/min.
 The tourniquet inflated pressure should not be affected by disturbances (excluding
external pressure changes) by more than +30 to -15 mmHg.
 The tourniquet inflated pressure should compensate for pressure changes due to
external pressure changes on the cuff within a few seconds
IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
21