3 Options

3.1

Settings

By pressing
(sub menu icon) you enter the sub menu. If no icon is pressed within 30
seconds, the main menu returns automatically.

3.1.1 Information

By pressing
(information icon) data regarding the unit will occur (manufacturer, model,
serial number and software version). Press the back icon to return to the sub menu or wait 30
seconds and the main menu will return automatically.

3.1.2 Timer

By pressing
(timer icon) time can be set, counting up or down. The alarm time can be set
by using the numeric display. ! The Timer may not be set during procedure.
When the settings are done press the "OK" icon and you will return to sub menu. If the new
setting data not is confirmed within 30 seconds, the new data will not be saved and the main
menu will automatically return.
By pressing "cancel data" the old timer settings will be valid and you will return to sub menu.

3.1.3 Self Test

By pressing
(self test icon) a self test will automatically be performed. When the self test is
done a new menu will occur with the text "Confirm self test OK". Confirm by pressing the
"OK" icon and the main menu will return. If a cuff is pressurized the user has to confirm every
step of the self test by pressing the screen as a part of the internal security mechanisms. If the self
test detects errors, it stops and the device returns to the main menu and shows errors. (Ref.
section 2.4)

3.1.4 Print

When pressing
is not connected". By pressing "Cancel" you will return to sub menu or wait 30 seconds and the
main menu will return automatically.
When the printer is connected and turned on the message "Printer is ready" will show and a print
icon is visible. Press the "Print" icon and the details from the last run will be printed.
If an error occurs, when the printer communicates with the tourniquet, the message "Error,
reconnect printer" will show.
No printer to be connected and used together with the Tourniquet during procedure!
During operation make sure to not touch USB port and patient at the same time.

3.1.5 Language

By pressing
(language icon) a menu with all default languages appears on screen. Choose
language and automatically return to sub menu. Chosen language will be default.
(print icon) without printer connected you will receive the message "Printer
IFU EN T800, Rev 09/2016-07-07, art.no. 8004007