Common Alarm Messages - HPM Tourniquet 800-20 Manual

Table of Contents

Advertisement

4.5

Common alarm messages

Signal definition
Safe State
Icon:
Instrument error
Icon
critical battery
low battery
Invalid pressure
High pressure
Low pressure
Critical low pressure
Critical high pressure
Inflating leakage
Inflated leakage
Alarms during startup test:
The following alarms after the power-up self-test will make the instrument in-operable
(instrument error).
Restart instrument, or call service!
Alarm message
Instrument error
System pressure failure
Pump failure
Pump leakage
Deflation valve failure
Cuff sensor failure
Cuff sensor failure
Distal block valve failure
Proximal block valve failure
The state when the block valve, the deflate valve and the pump all are
de-energized creating independent means to prevent the increase or
decrease of a cuff pressure. Monitoring of pressure values continues.
Make self test to clear this state.
Try to restart or contact service.
The battery capacity is less than enough for running one burst of high
priority alarm and switching the system in safe state after a serious
fault is detected.
The battery capacity can maintain monitoring and valve control but no
pump control (pressure control) for less than 10 minutes
If the measured pressure is invalid
If a pressure higher than 30 mmHg above Pset set value.
If pressure is lower than 15mmHg below Pset
If pressure is lower than Pset- 75 mmHg
If pressure is higher than 100 mmHg above Pset
If the set pressure value has not been reached within 20 seconds
The pump can not keep the set pressure in the cuff
IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
17