Recommended Test Equipment - Agilent Technologies 34980A Service Manual

Multifunction switch/measure unit
Hide thumbs Also See for 34980A:
Table of Contents

Advertisement

3
Calibration Procedures

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment
procedures is listed below. If the exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½-digit Digital
Multimeter to measure less accurate yet stable sources. The output value measured from
the source can be entered into the instrument as the target calibration value.
Application
Zero Calibration
DC Voltage
DC Current
Resistance
AC Voltage
AC Current
Frequency
Analog Output 34951A
Analog Output 34952A
Thermocouple
Reference Junction
34921A with 34921T Only
Relay contact resistance
All switch modules
[1] Optional test if not using relay cycle count.
48
Recommended Equipment
None 
Fluke 5700A
Fluke 5700A/ 5725A 
Fluke 5700A
Fluke 5700A/ 5725A
Fluke 5700A/ 5725A
Agilent 33220A
Internal DMM
Internal DMM
J Type Calibrated
Thermocouple 
Triple Point Cell
[1]
Agilent Y1131A
Accuracy Requirements
4 -terminal all copper short 
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
<1/5 instrument 24 hour spec
± 0.1 ×
C
± 0.001 resolution
34980A Service Guide