6.3.2 Operating steps for advanced settings
A.
Some material characteristics may be similar in terms of sensor attached and full
coverage sensitivity changes. Therefore, material characteristics can also be
determined by increasing the amplitude via advanced settings.
B.
In general, materials that require the use of advanced settings are mainly for high
viscosity characteristics (such as catchup, chocolate syrup) and two-level detection
(such as oil/water).
C. To connect to the sensor, click the "Continuous Monitoring" button to change it into
" ...... " on the advanced setting screen. When the material characteristic curve is
obtained, disconnect immediately (click the " ...... " button to change it into
"Continuous Monitoring").
* Please note that parameter modification and configuration can only be done in
disconnected state. To modify, click "Write Parameter Value" to save the parameters
to the sensor.
D. Operating steps:
1. Attach some material to the sensor head after the sensor is installed.
2. Click Advance to open the page.
3. Click the "Continuous Monitoring" button. When the material characteristic curve
is obtained, click the " ...... " button to exit.
4. Click "Record 1" to keep the characteristic curve when the sensor is attached to
the material on the screen, and record the amplitude peak (the value in the Amp
box in the upper right corner of the screen).
5. Add material so that it fully covers the probe of the sensor.
6. Click "Continuous Monitoring". In addition to the "Record 1" curve, the second
material characteristic curve appears on the screen, which is the curve when the
material fully covers the probe. Click again to exit.
7. Although the sensitivity is almost identical between the attached and fully covered
states, the amplitude peak of the signal is different. The amplitude peak of the
fully covered state is higher than that of the attached state.
8. If the amplitude peak of the fully covered state is 3300, the amplitude peak of the
attached state is 3000.Check the output port (OUTx) and "Amplitude
Determination Function" first, then drag the Y-axis determination line on the
screen to 3200 (or enter directly into the "Amplitude Action Point" field), and set
"Amplitude Hysteresis Point" to 100.
9. The recommended values in "Common Material Selection" can be used as the
sensitivity setting, including water, sugar-containing medium with low water
content or oil/powder.
10. Click "Write Parameter Value" in the lower right corner of the screen to write data.
A message pops up when succeeded.