Buffer Battery for Alarm in
the Event of Power Failure
A buffer battery is provided which allows the controller
to operate autonomously for approximately 12 hours in
the event of power failure.
The controller itself will recharge the battery keeping it
charged and ensuring availability when needed.
Figure 18. Battery Label
Figure 19. Buffer Batteries
Connection to External
Contacts
Ethernet Setup
Plug in the ethernet cable into the equipment's
RJ-45 port.
The appliance is factory setup with the following IP
address 192.168.0.127.
This IP address must be setup according to your
corporate network, so follow these steps:
In the homepage, touch first user menu icon
second service menu icon
password is asked. Enter 2810 PSW and confirm.
36 |
Touchscreen Controller Display Functioning
and
. Now, the service menu
2810
Access to LAN setup menu by touching
change IP and SUBNET address. Once entered the new
addresses confirm by "OK", and touch
and exit from the menu.
In order to check if the procedure has been successful,
enter the new IP address in a computer's browser
(advised browser: MOZILLA FIREFOX) connected to the
ethernet line – a detailed view of the equipment main
functions should appear:
CAUTION: In the service menu there are
parameters and configurations that guarantee
the proper functioning of the equipment.
Any unauthorized tampering may irreversibly
compromise the equipment and warranty will
automatically expire.
icon and
to record
Medical Devices -Touch Controller