Philips HeartStart Instructions For Use Manual page 51

Home defibrillator
Hide thumbs Also See for HeartStart:
Table of Contents

Advertisement

parameter
CPR Coaching
adult ventilation
instruction
CPR Coaching
infant/child ventilation
instruction
CPR Coaching
• 30:2 adult and
compression:ventilation
ratio
• 30:2 adult and
• 15:2 adult and
settings
Yes, No
Yes, No
30:2 adult and
30:2 infant/child
30:2 infant/
child
15:2 infant/child
15:2 infant/child
default
default description
Yes
Optional CPR Coaching includes
rescue breaths at the rate determined
by the CPR Coaching
compression:ventilation ratio for
adults when an adult pads cartridge is
installed.
NOTE: if this parameter is configured to
NO, CPR Coaching will always be
compressions-only when an adult pads
cartridge is installed.
Yes
Optional CPR Coaching includes
rescue breaths at the rate determined
by the CPR Coaching
compression:ventilation ratio for
infants and children when an infant/
child pads cartridge is installed.
NOTE: if this parameter is configured to
NO, CPR Coaching will always be
compressions-only when an infant/child
pads cartridge is installed.
If the user presses the i-button for
optional CPR Coaching during a
protocol pause or NSA pause, the
HeartStart provides coaching in basic
CPR for cycles of 30 compressions and
2 ventilations for adults, children, and
infants. Pauses begin and end with
compressions.
F
F-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

M5068a

Table of Contents