Download Print this page
laerdal Upright Oxygen Kit User Manual

laerdal Upright Oxygen Kit User Manual

Advertisement

Quick Links

User Guide
Upright
OXYGEN KIT (COMPLETE)
REUSABLE
REF Cat. no. 8561733 QTY 1 each
Spare parts
Cat. no
Description
846131
Oxygen Reservoir Bag and Tubing
All manuals and user guides at all-guides.com
User Guide
Intended Use
Laerdal Upright Resuscitator ("Upright") is a
self-inflating, manual, reusable resuscitator intended
for newborns and infants up to 10 kg body mass who
require respiratory support.
Upright can provide supplemental oxygen only when
used with the accessory Oxygen Kit.
The Oxygen Kit may be reused provided reprocessing
procedures (page 2) are followed between each
patient use. The kit (except for the Oxygen Reservoir
and the Oxygen Tube) must be cleaned and
disinfected before first use.
Using the Oxygen Kit
Assembly and disassembly of the Oxygen Kit is
illustrated on page 2.
Upright's pressure release valve can be operated as
normal when the Oxygen Kit is attached.
If any components are loose, then tighten or
reassemble the device and test in accordance with
page 2.
Delivered Oxygen Concentration
Measured under room temperature conditions:
20 ml tidal volume
Oxygen
@ 60 breaths/minute
flow
without
with
[l/min]
reservoir
reservoir
2
40%
46%
4
52%
67%
6
58%
89%
Important Product Information
Regulatory
2460
This product is in compliance with the essential
requirements of Council Directive 93/42/EEC as
ammended by Council Directive 2007/47/EC.
Technical Information
Meets ISO 10651-4:2002/EN ISO 10651-4:2009,
Lung ventilators – Particular requirements for
operator–powered resuscitators.
Operating temperature
-18 °C to 50 °C
Storage temperature
-40 °C to 60 °C
External dimensions of
Approximately
Upright when using the
72mm x 375mm x
Upright Oxygen kit
284mm
(with Newborn Mask size 1,
excluding tube)
Mass of Upright with
Approximately
Oxygen kit
288 grams
Materials
Hard plastic components: Polysulfone (PSU)
Soft plastic components: Silicone rubber
Oxygen Reservoir and Tubing: Polyvinyl chlorine (PVC)
Oxygen Reservoir Bag's Connector: Polycarbonate (PC)
Not made with natural rubber latex.
Safety When Using Oxygen
Build-up and transfer of high pressure to the
patient is prevented as excess oxygen is vented
to the atmosphere over the outlet valve.
If the reservoir bag remains flat during the whole
ventilation cycle, it is a a visual indication that no
or little supplemental oxygen is being provided.
When oxygen supply is insufficient, adequate
ventilation volume is ensured by intake of ambient
air through the Inlet valve.
In the presence of high oxygen concentrations
there is danger from smoking or naked flames. Oil
or grease must not be used with the resuscitator.
Warnings
This Oxygen Kit should only be used by persons
who have received sufficient training in its use.
Incorrect operation of the Oxygen Kit can be
hazardous.
Do not use the Oxygen Kit if you have any reason to
be concerned about its functionality.
150 ml tidal volume
@ 25 breaths /minute
without
with
reservoir
reservoir
52%
71%
63%
92%
73%
96%
Manufacturer Information
Global Warranty: See www.laerdal.com
Manufactured in China for:
Laerdal Medical AS, Tanke Svilandsgate 30
P.O. Box 377, 4002 Stavanger, Norway
Tel : +47 51 51 17 00
Fax: +47 51 52 35 57
Email: laerdal.norway@laerdal.no
Pending US and international design registration.
Laerdal® is a trademark or registered trademark
of Laerdal Medical AS.
© 2017 Laerdal Medical AS, All rights reserved.
Page 1

Advertisement

loading

Summary of Contents for laerdal Upright Oxygen Kit

  • Page 1 Laerdal® is a trademark or registered trademark Oxygen Reservoir and Tubing: Polyvinyl chlorine (PVC) of Laerdal Medical AS. Oxygen Reservoir Bag’s Connector: Polycarbonate (PC) Not made with natural rubber latex. © 2017 Laerdal Medical AS, All rights reserved. Page 1...
  • Page 2 Pull off Pull out using a blunt pair of tweezers For proper function, make sure to only use parts provided with the Upright Oxygen kit. Cleaning and Disinfection procedure Disassemble Clean Disinfect by one of the methods Dry and Inspect...