Pura Online Hygrometer User Manual
3.4.2
Configure Alarm Set-Point Limits
The following procedure is used to set limits
to which the alarm levels can be set (usually
after re-configuring the instrument's range
for Fahrenheit readings).
Figure
35
sequence.
1. Press the '
'
tECH
'.
2. Press the '
flash between '
3. Press the '' key four times and the display
will flash between '
4. Press the '
'
SU-L
'.
5. Press the '' key once to display a flashing
4 digit number representing the current
minimum alarm level setting.
setting for the ºC range is -100.0).
6. Use the '' and '' keys to set the required
new value (e.g. -148.0).
7. Press the '
The display will read '
8. Press the '' key once to display a flashing
4 digit number representing the current
maximum alarm level setting.
setting for the ºC range is 020.0)
9. Use the '' and '' keys to set the required
new value (e.g. 068.0).
10. Press the '
followed by the '
display.
Michell Instruments
shows
the
operational
P
' key once, the display will read
SET
' key once and the display will
ConF
' and '
PinP
ConF
' and '
GEnn
SET
' key once, the display will read
SET
' key to accept the new value.
SU-u
'.
SET
' key to accept the new value,
P
' key to return to the main
key
'.
'.
(The default
(The default
Figure 26
OPERATION
Main display
P
SET
x4
SET
SET
P
SET
Main display
Set-up alarm set-point limits