Valhalla Scientific 4314Ki Operation Manual page 27

Wide-range digital igniter tester
Table of Contents

Advertisement

LogOn2
Turns on continuous reading mode synonymous with "Screen? + RangeV?".
Note: if the device is already Logging, the LogOn2 command will return
"Invalid Command"
Syntax:
LogOn2<crlf>
Returns:
Screen data + the value of the range every 500ms.
Example for 2Ω Range:
Example for 20Ω Range:
Example for 200Ω Range:
Example for 2kΩ Range:
Example for 20kΩ Range:
Example for 200kΩ Range:
Example for 2MΩ Range:
LogOff
Stops continuous measurement mode.
Syntax:
LogOff<crlf>
Response:
<crlf>
Ohms?
Responds with the front panel reading formatted in ohms with multiplier.
No leading zeros.
Syntax:
Ohms?<crlf>
Returns:
Display data in Ohms with multiplier.
Example for a 1.1 ohm reading on:
2Ω Range:
20Ω Range:
200Ω Range:
2kΩ Range:
RangeN?
Responds with the range number from 1 to 8
Syntax:
RangeN?<crlf>
Returns:
R# (1-8)
R1
-
2 Ohms Range
R2
-
20 Ohms Range
R3
-
200 Ohms Range
R4
-
2k Ohms Range
COMMAND
COMMAND
QUERY
1.1000
1.100
1.10
0.0011k
QUERY
1.2345<crlf>2
12.345<crlf>20
123.45<crlf>200
1.2345<crlf>2k
12.345<crlf>20k
123.45<crlf>200k
1.2345<crlf>2M
R5
-
R6
-
R7
-
R8
-
25
20k Ohms Range
200k Ohms Range
2M Ohms Range
20M Ohms Range